12 - 51
12
Removing Disposables
1.
Disconnect the disposable tubing set from the catheter.
NOTE:
The disposable tubing set does not need to be disconnected from the syringe.
2.
Remove the syringe by twisting counterclockwise ¼ turn and pulling out.
NOTE:
If the syringe cannot be removed, rotate the injector head manual knob approximately two
revolutions in the forward direction, then repeat Step 2 above.
NOTE:
Once syringe has been removed, the piston automatically retracts to the start position (if
enabled).
3.
Discard fluid source containers and all used disposable items in accordance with facility pol-
icy.
Biological Contamination Hazard - Serious patient and/or worker injury or death may result.
•
Properly discard fluid source containers and disposable items after use (refer to
disposable label for specifics), or if there is any possibility that contamination may have
occurred.
•
Do not reuse single-use fluid sources. Refer to the fluid source manufacturers’
instructions for use.
WARNING
Summary of Contents for Certegra
Page 2: ......
Page 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Page 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Page 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Page 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Page 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Page 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Page 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Page 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Page 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Page 103: ......