10 - 41
10
Arming and Injecting
•
•
•
•
•
•
•
10.1 Add Volume Indicator
Whenever the Total Volume is greater than the Volume Remaining, the Home Screen provides
on-screen Add Volume indicators to communicate how much fluid should be added to perform the
protocol. Refer to Figure 10 - 1: Add Volume Indicators.
Figure 10 - 1: Add Volume Indicators
NOTE:
It is recommended that the operator resolve this condition prior to arming the injector by
either adding more fluid to the syringe or modifying the protocol.
10.2 Arm the Injector
The system can be armed from either the control room or the scan room.
10.2.1 Arm from the Control Room
1.
Select
Arm
on the display.
NOTE:
If the last piston movement was a reverse movement or a syringe required by the protocol is
not present, the system does not permit arming and communicates this to the operator.
2.
Perform the Check for Air Confirmation. Refer to Figure 10 - 2: Check for Air Confirmation.
Summary of Contents for Certegra
Page 2: ......
Page 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Page 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Page 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Page 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Page 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Page 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Page 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Page 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Page 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Page 103: ......