MEDRAD
®
Stellant CT Injection System with Certegra
®
Workstation
18 - 94
Guidance and manufacturer's declaration - electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below. The customer or user of the system should assure that it is used in
such an environment.
Immunity test
IEC 60601 Test Compliance Level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3Vrms from 150kHz to 80 MHz at 80% AM 1kHz
6Vrm, 80% AM 1kHz at ISM frequencies listed below:
WARNING:
Portable RF communications equipment
(including peripherals such as antenna cables and
external antennas) should be used no closer than 30
cm (12 inches) to any part of the injection system
unless a greater separation distance is required as
indicated by the equation. Otherwise, degradation of
the performance of this equipment could result.
Recommended separation distance
d
= 1.17
Radiated RF
IEC 61000-4-3
3Vrms from 80 MHz to 2.7 GHz at 80% AM 1kHz
and specific ISM bands listed below:
d
= 1.17
80 MHz to 800 MHz
d
= 2.33
800 MHz to 2.7 GHz
Where
p
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
a
should be
less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Non-ionizing Radiation Symbol (IEC TR
60878, 5140)
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromag-
netic site survey should be considered. If the measured field strength in the location in which the system is used exceeds the applicable RF compliance
level above, the system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Frequency
(MHz-ISM List)
Test Level
(Vrms)
1.8 - 2.0
6
3.5 - 4.0
6
5.3 - 5.4
6
6.765 - 6.795
6
7.0 - 7.3
6
10.1- 10.15
6
13.553 - 13.567
6
14.0 - 14.2
6
18.07 - 18.17
6
21.0 - 21.4
6
24.89 - 24.99
6
26.957 - 27.283
6
28.0 - 29.7
6
40.66 - 40.70
6
50.0 - 54.0
6
Frequency
(MHz)
Modulation
Type
Modulation
Frequency
Field
Strength
(Volts/meter)
385
Pulse
18 Hz
27
450
Pulse
18 Hz
28
710
Pulse
217 Hz
9
745
Pulse
217 Hz
9
780
Pulse
217 Hz
9
810
Pulse
18 Hz
28
870
Pulse
18 Hz
28
930
Pulse
18 Hz
28
1720
Pulse
217 Hz
28
1845
Pulse
217 Hz
28
1970
Pulse
217 Hz
28
2450
Pulse
217 Hz
28
5240
Pulse
217 Hz
9
5500
Pulse
217 Hz
9
5785
Pulse
217 Hz
9
Summary of Contents for Certegra
Page 2: ......
Page 4: ...MEDRAD Stellant CT Injection System with Certegra Workstation ii ...
Page 18: ...MEDRAD Stellant CT Injection System with Certegra Workstation 2 10 ...
Page 34: ...MEDRAD Stellant CT Injection System with Certegra Workstation 6 26 ...
Page 36: ...MEDRAD Stellant CT Injection System with Certegra Workstation 7 28 ...
Page 54: ...MEDRAD Stellant CT Injection System with Certegra Workstation 10 46 ...
Page 58: ...MEDRAD Stellant CT Injection System with Certegra Workstation 11 50 ...
Page 60: ...MEDRAD Stellant CT Injection System with Certegra Workstation 12 52 ...
Page 70: ...MEDRAD Stellant CT Injection System with Certegra Workstation 13 62 ...
Page 86: ...MEDRAD Stellant CT Injection System with Certegra Workstation 16 78 ...
Page 103: ......