07-19-61-244 October 2, 2009
HomeChoice
APD Systems Patient At-Home Guide
18-35
18. Troubleshooting
18.3.18 Temp Stabilizing
TEMP STABILIZING
Message Display:
Cause:
The temperature of the fluid in the heater bag measured by
the cycler is above 40°C.
Fluid delivery is not allowed at this time.
The room temperature combined with the heat from the
system (even with the bag heater off) is raising the solution
bag temperature too high to deliver safe fluids to you.
To Correct:
1. Press the power switch OFF.
2. Allow the system to cool for 10 to 20 minutes.
Bags containing less fluid cool in less time.
3. Press the power switch ON.
POWER RESTORED appears on the display screen.
4. Press
STOP
.
5. Press
GO
.
Therapy automatically resumes if the fluid temperature
falls below 40°C.
If the TEMP STABILIZING message continues to appear:
1. Press the power switch OFF.
2. Wait another 20 minutes.
3. Repeat Steps 3 through 5 above.
TEMP STABILIZING
Summary of Contents for HomeChoice
Page 12: ...Section 1 Glossary 1 Glossary...
Page 27: ...Section 2 User Assistance Information 2 User Assistance Information...
Page 30: ...Section 3 Warnings and Cautions 3 Warnings and Cautions...
Page 47: ...Section 4 Indications for Use 4 Indications for Use...
Page 50: ...Section 5 Description 5 Description...
Page 67: ...Section 6 Environmental Conditions 6 Environmental Conditions...
Page 70: ...Section 7 Setup and Check out 7 Setup and Check out...
Page 75: ...Section 8 Operating Instructions PRO Card and Modem 8 PRO Card and Modem...
Page 94: ...Section 9 Operating Instructions Change Program 9 Change Program...
Page 157: ...Section 10 Operating Instructions Make Adjustments 10 Make Adjustments...
Page 174: ...Section 11 Operating Instructions Prepare for Therapy 11 Prepare for Therapy...
Page 201: ...Section 12 Operating Instructions Perform Therapy 12 Perform Therapy...
Page 224: ...Section 13 Operating Instructions End Therapy 13 End Therapy...
Page 237: ...Section 14 Operating Instructions Effluent Sampling 14 Effluent Sampling...
Page 244: ...Section 15 Cleaning 15 Cleaning...
Page 248: ...15 Cleaning 15 4 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 249: ...Section 16 Maintenance 16 Maintenance...
Page 251: ...16 Maintenance 16 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 252: ...Section 17 Storage 17 Storage...
Page 254: ...17 Storage 17 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 255: ...Section 18 Troubleshooting 18 Troubleshooting...
Page 324: ...Section 19 Technical Data 19 Technical Data...
Page 345: ...Section 20 Quick Reference 20 Quick Reference...
Page 368: ...Index Index...