07-19-61-244 October 2, 2009
HomeChoice
APD Systems Patient At-Home Guide
18-19
18. Troubleshooting
18.3.8 Slow Flow Rate
SLOW FLOW
Display Message:
Cause:
Flow rate is very slow. A slow flow rate can reduce the Dwell
Time and decrease the amount of effective dialysis time.
There can be a partial kink on the specified line.
This is an Auto Restart Alarm.
To Correct:
1. Check the line shown in the display for:
–
Kinks
–
Closed clamps
–
Fibrin blockage
–
Disconnected solution bags
–
Empty solution bags
2. Correct the problem you found.
You do not need to press any buttons.
– OR –
If the alarm becomes a Manual Restart Alarm:
1. Press
STOP
to mute the alarm.
2. Check the line shown in the display for:
–
Kinks
–
Closed clamps
–
Fibrin blockage
–
Disconnected solution bags
–
Empty solution bags
3. Correct the problem you found.
4. Press
GO
to return to the therapy.
SLOW FLOW DRAIN
SLOW FLOW HEATER
SLOW FLOW PATIENT
SLOW FLOW SUPPLY
Summary of Contents for HomeChoice
Page 12: ...Section 1 Glossary 1 Glossary...
Page 27: ...Section 2 User Assistance Information 2 User Assistance Information...
Page 30: ...Section 3 Warnings and Cautions 3 Warnings and Cautions...
Page 47: ...Section 4 Indications for Use 4 Indications for Use...
Page 50: ...Section 5 Description 5 Description...
Page 67: ...Section 6 Environmental Conditions 6 Environmental Conditions...
Page 70: ...Section 7 Setup and Check out 7 Setup and Check out...
Page 75: ...Section 8 Operating Instructions PRO Card and Modem 8 PRO Card and Modem...
Page 94: ...Section 9 Operating Instructions Change Program 9 Change Program...
Page 157: ...Section 10 Operating Instructions Make Adjustments 10 Make Adjustments...
Page 174: ...Section 11 Operating Instructions Prepare for Therapy 11 Prepare for Therapy...
Page 201: ...Section 12 Operating Instructions Perform Therapy 12 Perform Therapy...
Page 224: ...Section 13 Operating Instructions End Therapy 13 End Therapy...
Page 237: ...Section 14 Operating Instructions Effluent Sampling 14 Effluent Sampling...
Page 244: ...Section 15 Cleaning 15 Cleaning...
Page 248: ...15 Cleaning 15 4 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 249: ...Section 16 Maintenance 16 Maintenance...
Page 251: ...16 Maintenance 16 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 252: ...Section 17 Storage 17 Storage...
Page 254: ...17 Storage 17 2 HomeChoice APD Systems Patient At Home Guide October 2 2009 07 19 61 244...
Page 255: ...Section 18 Troubleshooting 18 Troubleshooting...
Page 324: ...Section 19 Technical Data 19 Technical Data...
Page 345: ...Section 20 Quick Reference 20 Quick Reference...
Page 368: ...Index Index...