PREPARING THE MACHINE
| Attach the Heparin Syringe
It is important to correctly install the heparin syringe, it should be installed as follows;
1.
Insert the plastic collar of the syringe into the groove of
the pump
2.
NOTE:
To be able to use a syringe with a 13 mm inner
diameter, a heparin pump adapter must be installed in
the heparin pump
3.
Insert the plate on the end of the plunger into the groove
on the piston
4.
Turn the locking wheel upwards until resistance is felt
5.
Check the syringe is firmly in place by lightly pulling on
the plunger
1
2
3
4
5
HC10139 |