118
HYGIENIC MAINTENANCE I
Cleaning, decalcification and disinfection schedule
Frequency
Activity
Result
When more than 7 days passed since
last disinfection
Run a disinfection program before treatment
Disinfection
Every 1-3 months
1. Change the ultrafilter
2. Run a disinfection program
Disinfection
•
Do not perform more than 12 sodium hypochlorite based disinfections during the life cycle of the U9000
Ultrafilter.
•
Do not perform more than 8 sodium carbonate, e.g.
CleanCart A*
cartridge based disinfections
during the life
cycle of the U9000 Ultrafilter.
•
Disinfection using sodium carbonate e.g.
CleanCart A*
cartridge, should not be performed before periods when
the machine is inactive, e.g. storage over weekend.
•
The recommended process of heat disinfection using
CleanCart A*
cartridge should be followed by a heat
disinfection using
CleanCart C*
cartridge the same (working) day and should preferably be performed in the
middle of the working week.
*Not available in all markets
HC10139 |