117
HYGIENIC MAINTENANCE I
Cleaning, decalcification and disinfection schedule*
Frequency
Activity
Result
After each treatment
1. Wipe the outside of the dialysis machine with 70% ethanol or 60% isopropanol or1% sodium
hypochlorite. In case of cleaning with hypochlorite, wipe afterwards with water.
2. Rinse the outside and flush the inside of the pick-up tubes with water. Let them dry naturally.
3. If the prime bucket is used, wipe the inside and outside with maximum 10% sodium
hypochlorite.
Exterior cleaning
After each treatment, or at least once
per day
Run a disinfection program
Disinfection
After each treatment, if no disinfection
program is run
Run a descaling program
Descaling
At least after every 3rd treatment or at
high usage of the equipment at least
once per day
Run a heat disinfection program together with CleanCart C* cartridge or a heat disinfection
program with liquid citric acid
Decalcification
Disinfection
At least once every 7th treatment day
Cleaning using CleanCart A*
1. Run a heat CleanCart A* program
2. Run a heat CleanCart C* program or a heat disinfection program with liquid citric acid
3. Rinse the outside and flush the inside of the pick-up tubes with water. Let them dry naturally.
Cleaning using sodium hypochlorite
1. Run a heat CleanCart C* or a heat disinfection program with liquid citric acid
2. Run a chemical disinfection with sodium hypochlorite
3. Wipe the outside and flush the inside of the pick-up tubes with 70 % ethanol. Let them dry
naturally
Cleaning
Decalcification
Disinfection
*Not available in all markets
HC10139 |
Summary of Contents for AK 98
Page 1: ...AK 98 Machine Introduction Training Material ...
Page 14: ...14 THE OPERATOR S PANEL Touch Screen Layout HC10139 ...
Page 52: ...Treatment module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapters 4 8 ...
Page 101: ...Single needle module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapter 5 ...