SINGLE NEEDLE
| Start the Treatment
•
Once the treatment has been started, press the
flashing arterial pressure control
to centralize
the alarm limits around the current value.
•
The alarm limits for the venous pressure are
automatically set
once the treatment has started
to +350mmHg (high limit) and +150mmHg (low limit). If it is necessary to change the limits,
press the venous pressure control. Alternately the limits can also be changed by in the blood
menu low and/or high venous pressure limit.
•
The stroke volume that can be achieved is determined by the size and position of the venous
pressure alarm window, and the potential expansion volume provided by the expansion
chamber.