SAFETY
11 Safety check
The function test described below ensures that the OccluSense
®
-System works properly
and that the masticatory pressure distribution is displayed correctly. It is therefore
necessary that the function test is performed once a day.
First of all, a visual inspection for external damages of the device must be made.
If damages of the handheld, charging station or test sensor are detected, e.g. broken
parts of the sensor lid, the device must not be used.
In this case, see chapter 19 “Troubleshooting”.
12 Function test
To initiate the sensor test of the handheld, the handheld has to be connected to the app
first. Then tap “Next” and follow the instructions on the screen
(Figure 20-23)
.
The function test is performed with the supplied test sensor, following these five steps:
1. Place the test sensor in the correct position and close the sensor lid
2. Switch on the handheld and establish a connection with the iPad
3. Follow the instructions in the OccluSense
®
-iPad-App. The function test
is performed automatically
NOTE
A sensor test notification will pop up every day. However, you can proceed
•
without conducting the sensor test. If a test has not been performed for the last
5 days the app will disable any live mode or recording functionality until the
sensor test has been performed successfully (Figure 20-23).