4.2 System features
The OccluSense
®
-System serves as an aid for recording and depicting of occlusal
conditions by presenting the pressure conditions of the contact points of the masticatory
force distribution. It does not generate therapy recommendations automatically.
Furthermore, the system does not control the treatment/therapy and there is no inter-
vention in vital functions.
The occlusal conditions are recorded by the handheld device with inserted sensor. The
recorded raw data, generated during the occlusion test, are stored locally on the hand-
held device. Subsequently the handheld transmits this data via wireless interface to the
OccluSense
®
-iPad-App. The OccluSense
®
-iPad-App saves and displays the received data
as a recording.
14
Electrical
connection
inductive
power transmission
(galvanically isolated)
Electrical
connection
WLAN
iPad
Power supply
Charger
OccluSense
®
Sensor
OccluSense
®
Handheld