32
12. Press the balance ON/OFF key until the desired
day is displayed.
13. Press the balance ENTER key. [nn xx] will be
displayed.
14. Press the balance ON/OFF key until the desired
month is displayed.
15. Press the balance ENTER key. [Yr xx] will be
displayed.
16. Press the balance ON/OFF key until the desired
year is displayed.
17. Press the balance ENTER key. The display will
revert back to normal weight indication screen.
Resetting Variables of a Setra
Balance to Factory Default
This function is to reset the parameters inside the bal-
ance, which are not user accessible. The TIME, DATE,
and CALIBRATION are not affected.)
1.
Turn the green power switch on the furnace con-
trol panel to the OFF position.
2.
Press & hold the # 4 key while turning the green
power switch to the ON position. (Hidden Key 4
procedure). The red display will indicate [BAL
CAL]. Release the # 4 key when the NCAT
beeps once.
3.
Press & hold the balance ON/OFF key with a
small device such as straightened paper clip.
(Note balance control port holes in control panel
sketch.).
4.
Press & hold the balance ENTER key with a
small device such as straightened paper clip.
(Note balance control port holes in control panel
sketch.).
SETRA B
ALANCE