![Barnstead International Thermolyne 859 Series Service Manual Download Page 76](http://html1.mh-extra.com/html/barnstead-international/thermolyne-859-series/thermolyne-859-series_service-manual_3942385076.webp)
76
Elapsed Time: 30:12
Sample Weight:1093g
Weight Loss: 57.7g
Percent Loss: 5.27%
Temp Comp: 0.00%
Calib. Factor: 0.29%
===================
Calibrated Asphalt Ctnt
4.98%
===================
30 533 57.7 5.27
29 534 57.6 5.26
28 534 57.5 5.26
27 535 57.3 5.24
26 536 56.9 5.16
25 537 56.4 5.15
24 538 55.7 5.09
23 539 55.0 4.94
22 541 54.1 4.85
21 542 53.1 4.75
20 542 52.0 4.63
19 543 50.7 4.51
18 544 48.1 4.40
17 544 46.7 4.27
16 544 45.1 4.12
15 543 43.3 3.96
14 542 40.7 3.83
13 540 39.7 3.72
12 536 35.5 3.65
11 534 34.7 3.45
10 512 26.9 2.46
9 511 14.1 1.29
8 492 10.3 0.32
Example of an Ohaus balance configured incorrectly:
7 481 0.0 0.0
0.0 weight loss during the first 1-7 minutes, 0.00%
6 477 0.0 0.0 for the temperature compensation factor in the header
5 472 0.0 0.0 of the test.
4 469 0.0 0.0
3 465 0.0 0.0
2 455 0.0 0.0
1 451 0.0 0.0
-----:------:--------:------------
OHAUS balance had incorrect parameter settings
T: Temp: Wt. Loss: %Loss
causing this type of error. AZT was set to .5
----------------------------------
AL was set to 1.
Filter Set Pt: 750°C
AZT should have been set to OFF and AL should
Chamber Set Pt: 538°C
have been set to 3.
Tested by:______________
Mix Type: ______________
Sample ID:______________
Time: 10:03:32
Date: 10-13-01
E
RRORS
Summary of Contents for Thermolyne 859 Series
Page 1: ...NCAT Asphalt Content Furnace Service Manual LT1275X3 2 13 04...
Page 4: ...4...
Page 11: ...11 INSTALLATION...
Page 12: ...12 INSTALLATION...
Page 92: ...92...
Page 93: ...93...
Page 96: ...96...
Page 97: ...97...
Page 106: ...106...
Page 107: ...107...
Page 108: ...108...
Page 109: ...109...