To reset balance
1.
Power down the NCAT furnace using the green
power switch. Press and the hold the number
“4” key on the keypad. While holding down the
“4” key, power up the NCAT using the green
power switch. Hold the “4” key until you hear
the NCAT beep. Release the “4” key.
2.
Press and hold the ON button. Release when
“Menu” appears on the balance display. “Menu”
will automatically change to “Cal”.
3.
Press the MODE button. “User” will be dis-
played.
4.
Press the ON button. “Reset” will be displayed.
5.
Press the ON button. “Yes” will be displayed.
6.
Press the ON button to reset the user menu to
the OHAUS default values. You will know this
when this has been done when you hear a short
train of beeps.
7.
Press the MODE button. “AL” (Averaging Level)
will be displayed. Select this by pressing ON.
This will display “AL-1”. The correct value is
AL3. In order to change the averaging level,
continue pressing MODE until AL3 is displayed.
Press the ON to accept the new value. The
menu option “AL” will be displayed again.
8.
Press and release the MODE button until the
menu option “Auto-O” is displayed. Press the
ON button to select. Repeatedly press the
MODE button until ”Off” is displayed. Press the
ON button to accept this option.
9.
Press and release the MODE button until the
menu option “End User” is displayed. Press the
ON button to select. This will take you back to
the User Menu.
10. Press the MODE button. “Setup” will be dis-
played. Press the ON button to accept the
choice. Display will read, “Reset”.
OHAUS B
ALANCE
21
Summary of Contents for Thermolyne 859 Series
Page 1: ...NCAT Asphalt Content Furnace Service Manual LT1275X3 2 13 04...
Page 4: ...4...
Page 11: ...11 INSTALLATION...
Page 12: ...12 INSTALLATION...
Page 92: ...92...
Page 93: ...93...
Page 96: ...96...
Page 97: ...97...
Page 106: ...106...
Page 107: ...107...
Page 108: ...108...
Page 109: ...109...