115
- Chamber temperature should be +/- 10°C difference
between unit temperature and probe temperature at
500°C.
Back view of chamber section (859 & 945 series) with
cover removed. A 1/4” diameter hole is predrilled through
the stainless steel sheet metal but is not predrilled
through the chamber insulation material. A 1/4” diameter
hole must be drilled through the insulation material to
access the chamber.
Balance Calibration
Please refer to the “Balance” section of this manual for
complete calibration procedures.
M
AINTENANCE
10.5”
Type K Chamber
Thermocouple
L
i
Calibration
Port