R5914634 /01 MDSC-8527
50
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MDSC-8527 is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDSC-8527 should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the MDSC-8527 requires
continued operation
during power mains
interruptions, it is
recommended that the
MDSC-8527 be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Summary of Contents for MDSC-8527
Page 6: ...R5914634 01 MDSC 8527 6 Welcome ...
Page 12: ...R5914634 01 MDSC 8527 12 Welcome ...
Page 13: ...13 R5914634 01 MDSC 8527 Display installation 2 ...
Page 19: ...19 R5914634 01 MDSC 8527 Daily operation 3 ...
Page 39: ...39 R5914634 01 MDSC 8527 Troubleshooting 5 ...
Page 41: ...41 R5914634 01 MDSC 8527 Important information 6 ...
Page 62: ...R5914634 01 MDSC 8527 62 Important information ...
Page 63: ......
Page 64: ...www barco com R5914634 01 2022 03 17 Barco NV President Kennedypark 35 8500 Kortrijk Belgium ...