background image

T

Tv

Gn

G1

G0

Gn-1

Tm

SSI Data

Clock

Figure 8. 

SSI-compatible

 interface 

timing

T

Duration

 of clock signal: minimum 2 

μ

Sec = 500

kHz; maximum 13 

μ

Sec = 77 kHz

Tv

Delay 

time

 maximum 360 ns

Tm

Minimum 

time

 between last rising edge and

reloading of SSI, approximately 24 

μ

Sec

Gn

MSB (here Gray Code)

24-bit Transmission

G1 = second LSB; G0 = LSB

24+E Transmission

G1 = LSB; G0 = error bit

25-bit Transmission

G1 = second LSB, G0 = LSB

SSI Cable

The maximum baud rate for reliable data transmission depends on the cable length. The shielded 

connection

 cable is connected to the

sensor connector and the ground terminal of the control cabinet.

Cable Lengths

Cable Length

< 25 m

< 50 m

< 100 m

< 200 m

< 400 m

Baud Rate

< 500 kHz

< 400 kHz

< 300 kHz

< 200 kHz

< 100 kHz

Troubleshooting

 and Error Codes

When an error occurs, the error output Qs or Qp is set. The LCD displays an error message.

Multiple

 errors may exist simultaneously. In this case, both Qs and Qp are set. The LCD displays the 

respective

 error messages. The error

status is requested via the RS422 GSI command. For example, a voltage that is too low causes a counter error. As a result, the GSI

command reports 00000110.

Error Message

Error Output 

(Active

 Low)

Bit Response

Sequence

Meaning

Qs

Qp

BLINDING

Active

01000000

Internal error or ambient light too strong.

LAS. ERR.

Active

Active

10000000

Measurement laser faulty: repair or replace sensor.

LOW VOLT

Active

Active

00000100

Voltage too low or error in measurement of supply voltage.

NO VALUE

00000000

First measurement 

after

 switching on: sensor not ready. The message

disappears 

automatically

 

after

 300 ms when the sensor is ready.

PLL UNLOCKED

Active

Active

00000010

Counter error: repair or replace sensor.

Active

00100000

Temperature is out of acceptable range (below –10 °C or above +70 °C ).

OVERTEMP

Active

 (Laser OFF)

Active

 (Laser 

Off)

00101000

Operating

 temperature is too high (above +85 °C inside the housing). The

measurement laser switches 

off.

 Turn the sensor 

off.

 The sensor may

operate again 

after

 a cool-down period.

DIST (mm) >
Maximum

Active

00010000

No target in range or misaligned sensor.

CAUTION: Do Not Disassemble for Repair
This device contains no user-serviceable components. Do not 

attempt

 to disassemble for repair. A 

defective

 unit

must be returned to the manufacturer.

 

 

L-GAGE

®

 LT7 Long-Range Time-of-Flight Laser Sensor

16

www.bannerengineering.com - Tel: +1-763-544-3164

P/N 120244 Rev. C

Summary of Contents for L-GAGE LT7 Series

Page 1: ...use in personnel safety applications A sensor failure or malfunction can cause either an energized or de energized sensor output condition Models Models Sensing Mode Laser Class Cable1 Sensing Range2...

Page 2: ...aser pilot laser for alignment Both lasers are aimed at the identical target spot The laser beams are collimated to focus on a compact spot even at long sensing distances Diffuse Retroreflective 10 m...

Page 3: ...range 5 mm hysteresis QA analog Mode 1 rising positive slope full sensing range Offset 0 Unit mm Serial RS422 Password OFF Hysteresis Manually adjustable in 1 mm steps symmetrically around the switch...

Page 4: ...Limit 1 and Discrete Limit 2 to the analog output When copying discrete limits to analog Discrete Limit 1 and Discrete Limit 2 must be at least 300 mm apart or the sensor will not copy the limits The...

Page 5: ...rence 60825 1 Amend 2 IEC 2001 E section 8 2 Class 2 Visible Pilot Laser Lasers that emit visible radiation in the wavelength range from 400 nm to 700 nm where eye protection is normally afforded by a...

Page 6: ...gnment Aid The measuring sensing laser is located where the pilot laser spot is For fine adjustment use bracket SMBLT7 with the fine adjust accessory kit SMBLT7F for an angle of up to 3 in both X and...

Page 7: ...30 V dc H RX Black RS422 Receiver data SS1 Clock J NC Violet K TX Gray Pink RS422 Transmitter data SS1 Data L Q2 Red Blue Discrete Output 2 M GND Blue 0 V GND Installation Notes Some targets pose spec...

Page 8: ...ement laser stays ON alternating with the pilot laser The sensor proceeds to QuickSet on the Main menu Teaching the Outputs with QuickSet The received signal value is displayed as a bar graph the more...

Page 9: ...ASSWORD 1234 and the message PASSWORD OK 9 Press to save the password The sensor proceeds to QuickSet on the Main menu Accessing QuickSet Teach the current conditions to output Q1 and output Q2 with Q...

Page 10: ...to select the digit and move it to the left by one digit location 14 Press when the cursor flashes on the first digit on the left and the value is correct The sensor saves the setting and returns to...

Page 11: ...or to adjust it follow the steps below 1 Press to access the Offset 0 menu OFFSET 0 OFFSET 426 or OFFSET 0 OFFSET CLEAR is displayed OFFSET 0 returns the sensor to the default setting OFFSET 426 repr...

Page 12: ...intact 4 Press ESCAPE to return to Run mode DIST in 4839 The sensor returns to the Main menu with factory default settings intact Serial Communications Choose a RS422 compatible SSI1 10 compatible or...

Page 13: ...it command ASCII text Data Whole numbers ASCII text Note In ASCII text Command Data spaces and uppercase lowercase letters are ignored RS422 Cable The RS422 interface is a reliable serial interface in...

Page 14: ...put desired value ACK Input Offset Setting All Outputs Up to 12000 mm plus or minus or 48000 in plus or minus 4 IVL 0 1 ACK Enable Visible Laser 0 pilot laser OFF 1 pilot laser ON ISB 0 1 ACK Input St...

Page 15: ...d Do not insert a period or comma For example 12 000 mm is 12000 mm 480 00 in is 48000 Analog Output QA Commands and Their Meaning Only for diffuse sensor models Do not insert a period or comma For ex...

Page 16: ...er error As a result the GSI command reports 00000110 Error Message Error Output Active Low Bit Response Sequence Meaning Qs Qp BLINDING Active 01000000 Internal error or ambient light too strong LAS...

Page 17: ...t Mode Select Analog Output Slope Select diffuse models only and Output Limit Manual Adjust Serial Interface RS422 or SSI compatible Serial Measurement Speed RS422 2 9 ms at 57 6 kBaud SSI 1 4 ms SSI...

Page 18: ...y pk TX SSI Data rd bu Output 2 PNP bu Gnd LT7PIDQ bannerengineering com bannerengineering com Class 1 L GAGETM 42 0 mm 1 65 93 0 mm 3 66 95 0 mm 3 74 42 8 mm 1 68 42 0 mm 1 65 81 5 mm 3 21 5 75 mm 0...

Page 19: ...minum backing for use with LT7 Temperature 20 C to 50 C 4 F to 122 F Approximate size BRT 250 250 mm 250 mm BRT 540 540 mm 540 mm BRT 700 700 mm 700 mm Retroreflective Tape Model Reflectivity Factor M...

Page 20: ...R ANY INCIDENTAL CONSEQUENTIAL OR SPECIAL DAMAGES RESULTING FROM ANY PRODUCT DEFECT OR FROM THE USE OR INABILITY TO USE THE PRODUCT WHETHER ARISING IN CONTRACT OR WARRANTY STATUTE TORT STRICT LIABILIT...

Reviews: