2× Ø 10.5, 255 mm apart
(clearance hole for
screw heads)
Width 6 mm
23526b GB/2019-07
11 / 25
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2 Preparation
Carefully unpack the generator and accessories and inspect them for completeness
or possible transportation damage. If any damage or defects are found, these must be
immediately noti
fi
ed in writing to the transportation company and to the supplier.
Before operation, the generator should be allowed to stand for 2 hours at its operating
location so that it can adjust to the ambient conditions.
2.1
Scope of delivery
1 high-performance ultrasonic generator, see delivery slip
1 user instructions
Additional accessories according to order - see delivery slip
2.2
Installation / connection
• Place the generator on a
fi
rm, level and dry surface. In doing so,
- observe the maximum weight, depending on the housing size and number of
HF units, may be up to 10 kg,
- do not obstruct the air supply below and next to the generator, or the air outlet
(10 cm distance).
- guard against moisture and wetness - risk of electric shock.
• Installation in an enclosed construction is only permissible on the condition that it is
possible to dissipate the maximum possible thermal loss. Each HF unit can lose up
to a maximum of 200 W!
2.2.1 Optional mounting bracket TGW 1
The generator can be hung on a vertical surface by installing a mounting bracket on
the right or left side.