electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4 Use
4.1
Instructions for use
• Prior to switching on, it must be ensured that connected ultrasonic transducer
systems are completely covered with liquid.
• The speci
fi
ed
fi
lling level must be observed during use in tanks or similar containers.
• Tanks:
Freshly-
fi
lled bath
fl
uid or
fl
uid that has remained in the oscillating tank for a longer
period of time must be degassed prior to use.
- Remove basket and other accessories from the oscillating tank.
- Place lid on top.
- For degassing, start the ultrasound for about 30 min
Time need to be extended with acidic cleaning solutions.
• Immerse the instruments to be cleaned in the cleaning tank using the basket or
holder.
• VORTEX reactors and SONOBLOC tube reactors:
- Ensure that unit is completely
fi
lled.
- Ensure the unimpeded
fl
ow of the liquid medium during
fl
ow mode.
- If one of the red LEDs lights up or starts
fl
ickering after start-up, an operating
fault has occurred, see chapter 6.2.