4.2
General use
Step 1:
Check cable connections
Verify that all connecting cables and couplings are correctly connected.
Step 2:
Prepare the sonication
To do so, see the corresponding user instructions for tank baths, ultrasonic transducer
systems, or reactors.
Prepare the sonicating medium and check that the
fi
ll level is correct.
Step 3:
Start ultrasound operation
Press the mains switch.
The fan on the rear of the generator turns on.
When the power is switched on, all LEDs light up for approx. 4 seconds at the HF
unit. The "TEMPERATURE" LED
fl
ashes during this time.
Turn on the HF unit with the "ACTIVE" switch.
After a few seconds, when the power has been achieved, the "CONTROL" LED
(green) for the working HF units will be lit up.
If "CONTROL" is not triggered or if "TRANSDUCER" or "DRY" light up, there is
an error or incorrect adjustment (see chapter 6.2).
Step 4:
Stop ultrasound operation
In order to stop the power output, press the "ACTIVE" switch.
To turn the generator off, press the mains switch on the rear.
For additional care instructions, see chapter 5.