Indicator

Signi

fi

 cance

Fault / Noti

fi

 cation

Remedial action

CONTROL

(green)

TEMPERATURE

(red)

TRANSDUCER

(red)

DRY
(red)

ACTIVE 

(green)

-

-

-

-

Does not light 

up or only 

weakly

No mains voltage 
to HF unit

- Fuses
- Power transistors

Consult manufacturer

Lights up

-

-

-

-

Target output 
attained

Error-free operation

-

Flickers

-

-

-

-

Current output 
deviates

Due to the operation, 
the output is not 
continuously attained 
(movement in bath?)

-

-

-

-

-

-

No mains voltage

Power supply 
interrupted?

Check building's mains 
fuse

Grounded plug loose?

Connect plug correctly

Mains switch not set to 
ON?

Activate the mains 
switch

No output

Defective fuse?

Consult manufacturer

Too little output

Incorrect adjustment of 
the transducer?

Consult manufacturer

Idle motion
(HF unit blocked 
after 30 seconds)

Transducer not 
connected?

Connect transducer

Feed cable defective?

Check feed cable

Flashes 

or dark

-

-

-

-

Output reduced

Incorrect adjustment of 
the transducer/load?

Test transducer/load 
(with: "Test sheets 
for ultrasonic high-
performance transducer 
system") or contact 
the manufacturer

Flashes

Lights up

-

-

-

Flashes or  

dark

-

Flashes or 

fl

 ickers

-

-

Flashes 

or dark

-

-

Lights up

-

Flashes 

or dark

Lights up

-

-

-

Temperature-
speci

fi

 c output 

reduction

Is the temperature inside 
the generator too high?

If the ambient 
temperature is < 45°C, 
check the fan(s)

-

Incorrect adjustment of 
the transducer?

Consult manufacturer

-

-

Flashes

-

-

Idle motion
(HF unit blocked 
after 30 seconds)

Transducer not 
connected?

Connect transducer

once per second

Feed cable defective?

Check feed cable

Flashes quickly 

or 

fl

 ickers

HF voltage or 
current too high

Too many/few 
transducer systems?

Consult manufacturer

Lights up

Short circuit

Defective transducer or 
feed cable?

Check transducer 
and feed cable

Flickers

-

-

Lights up

-

Reactive power 
too high

Dry running or incorrect 
transducer adjustment?

-  Check level of 

fl

 uid

- Consult manufacturer

Flickers

-

Short period

Occasional occurrence 
not critical

-

Permanently
Reactive power 
too high

Limit state for dry 
running attained?

-  Check level of 

fl

 uid

- Consult manufacturer

Flickers

-

Flickers

Flickers

-

Problems with 
the ultrasonic 
transducers

Transducer frequency 
of generator and 
transducer different?

Consult manufacturer

Summary of Contents for SONOREX TECHNIK TG Series

Page 1: ...3 4 High performance ultrasonic generator series TG for the operation of ultrasonic cleaning tanks TM ultrasonic immersible transducers flat transducer plates or reactors valid for Compact housing TG...

Page 2: ...the translation and the original of this document the original version will take precedence BANDELIN accepts no responsibility or liability for damage caused by improper handling or usage contrary to...

Page 3: ...dling or usage contrary to the intended purpose In the event of unauthorised alterations modifications both the warranty claim and the CE conformity will become void If service is required please cont...

Page 4: ...11 2 1 Scope of delivery 11 2 2 Installation connection 11 2 2 1 Optional mounting bracket TGW 1 11 2 2 2 HF connection for ultrasonic transducer 12 2 2 3 Remote control connection 13 2 2 4 RS 232 in...

Page 5: ...adjustment TRANSDUCER 20 6 2 4 Short circuit and open circuit TRANSDUCER 21 6 2 5 Dry running identification DRY 21 6 3 Customer service 22 6 3 1 Repairs and service 22 6 3 2 Decontamination certific...

Page 6: ...ble transducers of the same or also different frequencies can be connected to a TG 3003 generator The HF units are programmed to the intended operating frequency by the manufacturer see label s on the...

Page 7: ...om the mains mains frequency 50 or 60 Hz into high frequency energy with a frequency of 25 or 40 kHz Thanks to the ultrasonic transducers that are connected to the ultrasound generator the high freque...

Page 8: ...f conformity can be requested from the manufacturer by providing the serial number 1 4 Technical data The ultrasonic generator is interference free and marked with a CE Safety EN 61010 1 EMC EN 61326...

Page 9: ...EN 61 010 1 Overvoltage category II Degree of contamination 1 Permissible ambient temperature 5 to 45 C Permissible relative humidity up to 31 C 80 Permissible relative humidity up to 40 C 50 Condens...

Page 10: ...forbidden to connect any type of power or voltage sources to the generator s signal inputs or outputs If the unit is passed on to others the user instructions and the safety instructions must also be...

Page 11: ...ormance ultrasonic generator see delivery slip 1 user instructions Additional accessories according to order see delivery slip 2 2 Installation connection Place the generator on a firm level and dry s...

Page 12: ...the load The factory set data for the HF unit is affixed next to the HF connection socket HF output socket Connect the installed ultrasonic transducers cleaning tanks TM immersible transducers flat tr...

Page 13: ...elin com Note Multiple ultrasonic transducers of the same frequency can be connected to one HF unit up to the specified output on the type plate until the sum of the individual outputs is equal to the...

Page 14: ...rror TEMPERATURE red Flashes if the power is reduced due to excessively high temperatures inside the generator TRANSDUCER red Lights up permanently in the case of a short circuit Defective transducer...

Page 15: ...he oscillating tank for a longer period of time must be degassed prior to use Remove basket and other accessories from the oscillating tank Place lid on top For degassing start the ultrasound for abou...

Page 16: ...ess the mains switch The fan on the rear of the generator turns on When the power is switched on all LEDs light up for approx 4 seconds at the HF unit The TEMPERATURE LED flashes during this time Turn...

Page 17: ...er No guarantee is provided for damage caused by the use of unsuitable disinfection agents or detergents 5 1 Cleaning and care Fans cooling of the generator The generator must be cleaned regularly Eve...

Page 18: ...enerator functions are controlled by the program At the beginning of troubleshooting the mains switch should be turned off After approx 6 seconds it should be turned on again In the process the LEDs o...

Page 19: ...r load with Test sheets for ultrasonic high performance transducer system or contact the manufacturer Flashes Lights up Flashes or dark Flashes or flickers Flashes or dark Lights up Flashes or dark Li...

Page 20: ...th the cooling element temperature creates a working state in which the HF unit continues to work safely with a reduced output allowing work to continue Once the excessive temperature has been resolve...

Page 21: ...of the dropping of the liquid level in the cleaning tank If the liquid level above the surface of a connected ultrasonic transducer drops below 5 mm the ultrasonic output is considerably reduced in o...

Page 22: ...rlin Repair service E mail Phone 49 0 30 768 80 13 info bandelin com Fax 49 0 30 76 88 02 00 13 In the event of returns the General Terms and Conditions for Delivery and Payment of BANDELIN electronic...

Page 23: ...ntrol with timer BANDELIN order No 8466 or the FS 7 remote control cable BANDELIN order No 8468 Simple on off function via relay switch SPS etc This can be done by inserting the 5 pole plug of the res...

Page 24: ...y utilise the ultrasound Additional information may be obtained from our supplier our sales representatives or from our website No obligation telephone consultation Website 49 0 30 768 80 280 www dr s...

Page 25: ...Note The instructions for use in this and other languages as well as further information can be found on the enclosed CD...

Reviews: