24 / 42
51506-002 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.3.1
Mounting and dismounting of the probes
Probes
• are screwed onto the standard or booster horns,
• transmit ultrasonic power into the medium to be sonicated,
• are made of high-strength titanium alloy.
Caution! Probes are sensitive to shock.
Before mounting or dismounting the probes, the ultrasonic generator must be turned o
ff
and
the ultrasonic transducer must be disconnected from the ultrasonic generator.
The mounting surfaces
4
must be thoroughly cleaned so
that the amplitude and/or power can be transmitted to
the medium.
If the mounting surfaces are not cleaned, the probe,
standard horn or booster horn could be destroyed as
a result.
The tool required for mounting/ dismounting is included in the scope of supply.
4 The
mounting surfaces are the contact surfaces between the individual accessory parts, e.g. between the standard or booster horn and
the probe.