Ultrasonic transducer
Standard and/or booster horn
Vigorously turn sickle
spanner HS 40/42 long,
in the opposite direction
27 / 42
51506-002 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.4
Dismounting and mounting of the standard and booster horns
Standard and/or booster horns
• are screwed onto the respective ultrasonic transducer,
• are made of high-strength titanium alloy.
• transmit the oscillations to the probe,
• enhance the
amplitude thanks to their geometry.
At the time of delivery, standard horn SH 70 G and booster horn SH 213 G, are
fi
rmly mounted
onto the HD 2070.2 and HD 2200.2 respectively, on the ultrasonic transducer.
Before mounting the standard or booster horn, the ultrasonic generator must be turned o
ff
and
the ultrasonic transducer must be disconnected from the ultrasonic generator.
Release
the
fi
rmly-mounted standard and/or booster horn SH 70 G, SH 213 G or SH 3425
from the ultrasonic transducer UW 3100, UW 2070 or UW 2200, respectively. To do so, lay
the ultrasonic transducer on a
fi
rm base (non-slip, if possible).
Video available under:
https://www.youtube.com/watch?v=NXbGc6nAb5c
Disassembly
Place one sickle spanner on the ultrasonic transducer and the other sickle spanner on the
standard or booster horn. To dismount, press the
fi
rst sickle spanner against the
fi
rm base,
press the other sickle spanner