4
/
Preparation for operation
12/36
51846-000 en/2022-07
Fig. 2
Loosen (A) and tighten (B) the probe
A video on loosening and attaching the probe is available:
https://www.youtube.com/watch?v=hjTC_cvO4kQ
4.3 Mounting/removing standard horn or booster horn
NOTICE
Damage to the ultrasonic converter
If the ultrasonic oscillating system in the housing of the ultrasonic converter
twists, the ultrasonic converter will be damaged.
– Always use the sickle spanner to secure the standard horn or booster horn.
– Never hold the ultrasonic converter by the housing when tightening or loos-
ening the probe.
Requirement
• The ultrasonic converter is disconnected from the ultrasonic generator.
• Tools required: Sickle spanner for ultrasonic converter and standard horns or
booster horns (supplied tool set).
Procedure
1. Place a sickle spanner on the ultrasonic converter. Place the ultrasonic con-
verter on a fixed tabletop and support the sickle spanner on the tabletop.
2. Place a sickle spanner on the standard horn or booster horn. Loosen the
standard horn or booster horn.
20 mm
20 mm
B
A