24 / 38
51406c GB/2019-07
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.3
Selection of suitable probe
Detailed information on the individual probes can be found in the separate product information.
Probe
MS 1.5
MS 2.0
MS 2.5
Tip diameter
1.5 mm
2.0 mm
2.5 mm
Sonication volume
0.1 - 10 ml
0.25 - 20 ml
0.5 - 25 ml
Amplitude
65
μ
m
ss
70
μ
m
ss
72
μ
m
ss
Maximum permissible
amplitude setting
90 %
100 %
100 %
4.3.1
Mounting and dismounting of the probes
To install or remove the probes, 2 spanners of type SW 10 (8 x 10) are needed. The spanners are
included in the scope of delivery of the ultrasonic homogenizer or
are supplied together with the probe.
Do not mechanically burden the probes during installation
danger of bending.
Before
mounting the probes, the HF generator must be switched off or the ultrasonic converter
separated from the HF generator.
Work steps for disassembly (see illustration)
Place the ultrasonic converter on a level surface, e.g. tabletop.
Push the spanner with width 10 at the coupling piece into the guide grooves.
Turn the ultrasonic converter and spanner to the left so that the other end of the spanner lies
on the surface and the coupling piece can no longer be turned to the left.
Now push the spanner with width 8 into the guide grooves of the probe.
Then unscrew the probe. In doing so,
fi
rmly hold the spanner at the coupling
piece.