Monitor display
Error
No.
Signi
fi
cance
Cause / Measure
Remedial action
Error !!!
003
please
contact service
please
contact service
003
No
power output
Device fault
Repairs only
through Customer
Service
-
Send in homogenizer for
repair
since fault is severe
Error !!!
011
please
check the probe
002
Frequency setting not
possible
Ultrasonic
transducer or
probe
faulty
. Replace
parts and check
function (conduct
a “probe check” or
“scan frequency”)
If no success
Repairs only
through Customer
Service.
-
Check the connection to the
probe.
Is the cable properly plugged
in and af
fi
xed? Is the
probe
correctly screwed on? Check
for erosion on the
probe.
-
Once error recti
fi
ed,
conduct
the “probe check” or “scan
frequency” function (see
chapter 6.2).
01
1
No response signal
from the ultrasonic
transducer
012
No
resonance
frequency found
Error !!!
014
max. temperature
exceeded
014
Permissible internal
operating temperature
exceeded. Ultrasonic
homogenizer turns of
f.
Allow the ultrasonic
homogenizer to
cool down.
-
W
ait until the
HF generator
has suf
fi
ciently
cooled.
The
operating temperature must be
< 50 °C.