14 / 46
51346-007 GB/2020-08
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2
Preparation
Carefully unpack the ultrasonic bath and accessories and inspect them for completeness
or possible transportation damages. If any damages or defects are found, these are to be
immediately noti
fi
ed in writing to the transportation company and to the supplier.
Before startup, the ultrasonic bath is to be left to stand at its operating location for 2 hours
so that it may adapt to the ambient conditions.
2.1
Scope of delivery
1 × ultrasonic bath with integrated cooling unit (order No.: 3500) including:
1 ×
beaker holder PH 255-11
1 ×
lid
D 255 G
20 × inset beakers SD 01.2
1 ×
drain hose ¼"
1 ×
mains cable
2 ×
rubber stoppers (as transport safety devices)
1 ×
TICKOPUR TR 3
250 ml
1 Instruction
manual
Additional accessories according to order – see delivery note
2.2
Set-up / assembly
• Place the ultrasonic bath on a
fi
rm, level, dry surface, on a location away from direct
sunlight if possible. In the process
- Observe the maximum weight of the ultrasonic bath, including
fl
uid.
For net weight, see technical data chapter 1.4.
- Observe the minimum distances for the incoming and outgoing air.
At least 5 cm on the side and at least 10 cm on the rear.
- Guard against moisture and wetness - risk of electric shock.
• The ultrasonic bath must be positioned in such a way that disconnection from the power
supply is easily possible.
5 cm
5 cm
10
cm