1.4.2
Interfaces
The SONOCOOL ultrasonic bath features two interfaces on its rear side which may be
optionally used by the operator.
Note:
Implementation of the technical programs required for communication is the responsibility
of the operator and is not supported by the manufacturer. The manufacturer only
guarantees the proper functioning of the interfaces.
Commercially-available cables may be used for the communication.
It is possible to use both interfaces simultaneously.
1.4.2.1 D-Sub 15 - Pinning
15-pin D-Sub interface for remote control, e.g. using an CLP.
SubD
pin
Name of signal ML-pin Remark
1
PE
1
Ground potential connected to GND
2
TxD
1
3
SCI TxD
3
RxD
1
5
SCI RxD
4
--
7
Not connected!
5
GND
9
Combined ground potential for all signals
6
PWR_down1
11
Ultrasound disconnection during bridging
2
7
PWR_down0
13
8
--
15
Not connected!
9
--
2
10
--
4
11
\M_HF_on
6
Signal contact
3
for
ultrasonic emission
open against CLP_COMMON
12
\M_HF_on
8
closed against CLP_COMMON
13
CLP_COMMON
10
Combined contact for messages
14
\M_fail
12
Signal contact
3
for
errors
open against CLP_COMMON
15
M_fail
14
closed against CLP_COMMON
1 Optional
2
Switching with potential-free contact. Internal connection with 12 k
Ω
at +5V. Bridging voltage < 1V (e.g. with optocoupler).
Caution: Switching only with potential-free contact relay or optocoupler.
3
Contact may be loaded with 60V, 0.5A, maximum 10VA against CLP_COMMON