1 Power control knob ( power selection switch)
2 Green control light (ultrasound in operation)
3 Turning knob for ultrasound ON / OFF with preset time
1
2
3
12 / 32
6366-006 GB/2020-09
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2.3
Start-up
Thoroughly rinse the ultrasonic bath's oscillating tank with water before its
fi
rst use.
Note:
In order to protect the surface during transport and storage, all outer surfaces (and also
the inner walls of the oscillating tank) are covered with an oily preservative. This should be
removed with a suitable cleanser before
fi
rst use, see chapter 5.
Implant boxes:
Prior to initial usage and when reprocessing, the implant boxes are to be prepared according
to the instructions attached!
Verify that the operating button on the right-hand side is in the "o
ff
" position, then connect the
ultrasonic bath to the mains (grounded socket).
Conduct function test – brie
fl
y plug in the ultrasound (maximum of 1 to 2 seconds), a hissing
noise should be heard. Then switch device o
ff
again.
If applicable, hang accessories in the ultrasonic bath and place lid on top.
3 Operation
3.1
Operating elements
The ultrasound and the power are operated from the front:
3.1.1
Power selection switch
The power is controlled through the integrated power selection switch.
Adjustable power
settings
20% P
40% P
60% P
80% P
100% P
= 40 W
= 80 W
= 120 W
= 160 W
= 200 W
HF power
max. 200
W
e
ff
Note
• For
foil tests, for degassing and cleaning, etc., the power selection switch must be set to
100% P.