4.2
General application
Step 1:
Fill oscillating tank
The oscillating tank is
fi
lled with water and a suitable preparation to reduce the surface tension,
see chapter 7.3.
Indirect sonication
When using aggressive acidic liquids (e.g., sulfuric acid, hydrochloric acid), we recommend using
alkaline contact liquid such as 5% TICKOPUR R 33.
Fill 1/3 of oscillating tank with water.
Add
surfactant.
Fill carefully, avoid as much as possible
the formation of foam.
The oscillating tank must be
fi
lled
contingent upon the inset beaker since
inset beakers displace the contact liquid.
Direct sonication
Fill 1/3 of oscillating tank with water.
Add dosed preparation to the oscillating
tank. See appendix for dosage
information.
Fill carefully up to the
fi
lling level mark,
avoid as much as possible the formation
of foam.
Step 2:
Degassing the
fl
uid
Freshly-
fi
lled bath
fl
uid or
fl
uid that has remained in the oscillating tank for a longer period of time
must be degassed prior to use. See also chapter 4.3.1.
Remove basket and other accessories from the oscillating tank.
Place lid on top.
Set the power selection switch to 100% P.
Switch ultrasound on for 10 min, see chapter 3.1.2