2/3
Oscillating tank with
inset basket
2/3
Oscillating tank with box
holder and implant box
17 / 32
6366-006 GB/2020-09
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Step 3:
Insert objects to be treated
For optimal di
ff
usion of the ultrasound and protection of the oscillating tank, nothing should be in
contact with the bottom of the bath during sonication. Implant boxes and other inset beakers must
be stored in the supplied box holders or similar accessories. Suitable baskets are to be used for
placement of objects to be cleaned, see chapter 7.
Before every sonication it is necessary to check whether the sonication
fl
uid needs to be cleaned
or replaced.
Indirect sonication
Hang
the
box holder in the oscillating tank and
fi
ll it with implant
boxes/inset beakers.
Fill box/inset beaker, in doing so please note:
•
It is possible to treat multiple sample vessels with di
ff
erent
fl
uids
at the same time.
•
When treating small quantities of combustible
fl
uids, the country-speci
fi
c guidelines/
regulations that are currently in e
ff
ect are to be observed.
Place the box/inset beaker on the holder, in doing so please note:
•
Immersion depth of the box is min. 2 cm.
• Control
fi
ll level ( contact liquid).
• Remove
air bubbles below the box.
Direct sonication
Place the objects to be cleaned in the appropriate accessories, in doing so
please note:
•
Evenly distribute parts, do not stack them.
•
An overloading of the basket or inset beaker reduces the ultrasound
e
ff
ect (the ultrasound is absorbed).
•
Place the more heavily soiled side facing downward.
•
Parts with joints are to be fully opened before placing inside.
•
Delicate parts are not to come into contact with one another - for positioning, use special
accessories such as silicone knob mats if necessary, see chapter 7.
•
Due to the design, the ultrasound e
ff
ect is weaker on the outlet side. Heavily contaminated
objects should not be placed in the basket over the outlet.
Hang the inset basket with the goods to be treated or place the basket holder in the oscillating
tank and the inset basket on the basket holder.
Verify that the objects to be treated are fully covered by
fl
uid.
With every object inserted, the
fi
ll level is to be monitored.