BG 800 300 BE / A (0105) TPG 300
29
4.7.7 »Filter« group
In the event of fast varying measurement signals, the measured
values can be filtered to stabilize both, the display and the
switching functions.
a) »Filter assignment« function
You can set the filter separately for each individual measuring
circuit.
1 2 3 4 A B
A1
A2
B1
B2
0
10
The cycle depends on the plug-in boards installed.
b) »Filter time constant« function
Three filter time constants are available.
In the case of signal fluctuations, a faster filter can
cause 'fluttering' of switching functions.
1 2 3 4 A B
A1
0
10
Display
Filter
Time constant
FI 1
FI 2
FI 3
fast
medium (default)
slow
16 ms
160 ms
1.6 s
Any modification is made immediately.
The analog outputs are not affected.
4.7.8 »Interface« group
a) »Baud rate« function
1 2 3 4 A B
0
10
Display
Baud rate
bd 3
bd 1
bd 2
bd 4
bd 9
300
1200
2400
4800
9600 (default)
The baud rates for the TPG 300 and any interfaced
computer must be the same.