BG 800 300 BE / A (0105) TPG 300
57
Receive :
xxxx <CR><LF>
³
ÀÄÄ
x = 0000
Ä>
no error
1000
Ä>
TPG error (see display on the front panel)
0100
Ä>
NO HWR
Hardware not installed
0010
Ä>
PAR
Invalid parameter
0001
Ä>
SYN
Syntax error
The ERROR word is erased as it is read out. It is automatically reset if the error persists.
Example
"Send (S)" and "Receive (E)" are referenced to the host.
S:
E:
S:
E:
TID <CR> [<LF>]
<ACK> <CR> <LF>
<ENQ>
PI 300DL, PE 300-10, IF 300 <CR> <LF>
Request for TPG identification
Positive acknowledgment
Inquiry
Plug-in board types
S:
E:
S:
E:
SEN <CR> [<LF>]
<ACK> <CR> <LF>
<ENQ>
3, 3, 1, 0 <CR> <LF>
Request for sensor status
Positive acknowledgment
Inquiry
Sensor status
S:
E:
S:
E:
SPB <CR> [<LF>]
<ACK> <CR> <LF>
<ENQ>
1.0E-11, 9.0E-11, 0 <CR> <LF>
Request for parameter of switching function B
Positive acknowledgment
Inquiry
Threshold values and assignment
S:
E:
SPB, 6.8E-3, 9.8E-3, 2 <CR> [<LF>]
<ACK> <CR> <LF>
Modify threshold values of switching function B
Positive acknowledgment