BG 800 300 BE / A (0105) TPG 300
40
Problem
Possible cause
Correction
Unit cannot be locked
Code 99 19 activated
1. Pull the measurement plug-in boards approx.
1 cm out of the slots A and B
2)
2. Change the code in »set up« mode
3. Reinstall the measurement plug-in boards
2)
Code forgotten
---
1. Pull the measurement plug-in boards approx.
1 cm out of the slots A and B
2)
2. Select the code in »set up« mode
3. Read out the code
4. Reinstall the measurement plug-in boards
2)
2)
Please read the instructions in 3.2.2 before performing any manipulations on the plug-in boards.
6.2.3 Failures
Problem
Possible cause
Correction
No display appears when the unit is
switched on
No mains power
Check mains
Fuse defective
Replace fuse (
→
section 6.2.4)
Individual LEDs, or segments of the 7-
segment or bar graph display do not light
Display / microprocessor defective