389D001 Rev G
58 of 94
CHECKED
DRAWN
DATE
DATE
TITLE
THIS DRAWING AND ITS CONTENTS ARE THE
PROPERTY OF THE BAKER COMPANY. THE
INFORMATION CONTAINED HERE ON MAY NOT BE
USED OR COPIED IN ANY MANNER WITHOUT THE
WRITTEN PERMISSION OF THE BAKER COMPANY.
389E514
W.Day
LADDER SCHEMATIC
Protective conductor terminal
!
DENOTES WIRE NO. 54, 22 A.W.G.
ALL WIRES 16 A.W.G. UNLESS OTHERWISE NOTED
DENOTES CANOPY TERMINAL BLOCK, TERMINALS 3 OR 4
LEGEND
Caution(Refer to
Accompanying documents)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
A
SGX04-INT, W/Cabinet Setpoint Monitor
FOR
DWG NO.
TB-3,4
22
54 BL
SHEET
11/13/14
T.Martin
11/13/14
A
ORIGINAL RELEASE
DATE
DRN
APP-BY
APP-DATE
ALL ITEMS MUST BE
RoHS COMPLIANT
161 Gatehouse Road Sanford, ME 04073
Ph: 207-324-8773 / web: www.bakerco.com
2 OF 3
7
BK
RD
BK
RD
2
1
3
BN
18
BL
18
TB 1, #
4,5 o
r 6
TB 1, #
1,2 o
r 3
RD
BK
BK
RD
WT
WT
L+ N-
+ -
Output DC
24V 0.5A
DC
OK
Input AC
100-240V
PS1
B K
+
V
-
V
NO
1
C1
NC
1
NO
2
C2
NC
2
RD
B K
RD
RD
PM1
A 1
A 2
CR
1
14
11
12
B K
RD
RD
1
TB1, #9: +24V DC
TB1, #10: - 24V COMMON
TB1, #11: PRESSURE OK; 24VDC WHEN PRESSURE IN SAFE CONDITION
PM1: PRESSURE OK; RELAYS CLOSED WHEN IN A SAFE CONDITION
CR1 & CR2: PRESSURE OK; ON WHEN PRESSURE OK
8
9
PR
E
S
S
U
R
E
O
K
WT
A1
A2
CR
2
14
11
12
WT
+12V
SPARE
GND
J32
2
3
NO
C
1 2
1
3
FEMALE
REAR VIEW
1
3 2
MALE
REAR VIEW
BELDEN 22/4 CABLE
BK
RD
BK
BELDEN 22/2 CABLE
TO J32 ON THE
CONTROL BOARD
CUSTOMER ALARM CONNECTION
1 2
1
FEMALE
REAR VIEW
2 1
MALE
REAR VIEW
2
1
RD
BK
BELDEN 22/2 CABLE
CUSTOMER CONNECTION
FOR ANALOG PRESSURE
OUTPUT
1
NC
RD
WT
BK
NO
C
NC
RD
WT
BK
4/20
4/2
0
+
-
RD
BK
WT
BK
4
2
1
4
3
FEMALE
REAR VIEW
2 1
4 3
MALE
REAR VIEW
WT
BELDEN 22/4 CABLE
BELDEN 22/4 CABLE
BK
WT
REMOTE CONNECTION OPTION
Summary of Contents for SterilGARD SG404
Page 60: ...389D001 Rev G 60 of 94...
Page 74: ...389D001 Rev G 74 of 94 Auxiliary Blower Switch Status Wiring Option...
Page 75: ...389D001 Rev G 75 of 94 Auxiliary Cabinet Monitoring Wiring Option...
Page 76: ...389D001 Rev G 76 of 94 FlexAIR Canopy Exhaust Connection Wiring...
Page 77: ...389D001 Rev G 77 of 94...
Page 78: ...389D001 Rev G 78 of 94...
Page 79: ...389D001 Rev G 79 of 94...
Page 80: ...389D001 Rev G 80 of 94...
Page 81: ...389D001 Rev G 81 of 94...
Page 82: ...389D001 Rev G 82 of 94...
Page 83: ...389D001 Rev G 83 of 94...
Page 84: ...389D001 Rev G 84 of 94...
Page 85: ...389D001 Rev G 85 of 94...
Page 86: ...389D001 Rev G 86 of 94...
Page 87: ...389D001 Rev G 87 of 94...
Page 88: ...389D001 Rev G 88 of 94...
Page 89: ...389D001 Rev G 89 of 94...
Page 90: ...389D001 Rev G 90 of 94...
Page 91: ...389D001 Rev G 91 of 94...
Page 92: ...389D001 Rev G 92 of 94...
Page 93: ...389D001 Rev G 93 of 94...