389D001 Rev G
15 of 94
III - CERTIFICATION
Field Certification Procedures
The following Field Certification Procedures shall be conducted by a trained biosafety cabinet technician.
Cabinet Airflow Verification
The following calibrated test instruments and tools are required to perform this field certification.
Direct Inflow Measurement Device (For Primary Inflow Method) (optional)
Thermal Anemometer (For Secondary Inflow Method and Downflow air velocity measurement)
Aerosol Photometer and Laskin Nozzle Generator
Smoke Visualization Device (Smoke tubes)
Amperage meter
1/4” to 7/16” Standard (English) ratchet or socket style tools. All bolts are Imperial, not Metric
Adjustable Wrench
Good quality adhesive tape
Step ladder
Tape measure with Imperial measurements
Pocket calculator
Flat head and cross head screwdrivers
The SterilGARD Biosafety Cabinet is shipped with a protective laminate sheet covering the exhaust HEPA
filter. Removal of this sheet shall be done by the certifier, NOT THE INSTALLER, because of potential
damage during installation. If this protective sheet is not present after installation, inspection of the exhaust
HEPA filters is strongly recommended. Filter replacement maybe necessary if damage found cannot be
repaired.
Step 1
: If equipped with FlexAIR Canopy Exhaust Connection, be sure to have the exhaust air turned on and at
the specified exhaust flow range before beginning the cabinet airflow validation procedure. (See Baker Document
BAKD066 for exhaust airflow requirements)
To meet manufacturer’s specifications, all methods and acceptance
criteria must be followed. No alternative or secondary methods for
measuring cabinet airflow other than what is stated in this service manual
and the cabinet data plate shall be used to validate the performance of this
biosafety cabinet.
Summary of Contents for SterilGARD SG404
Page 60: ...389D001 Rev G 60 of 94...
Page 74: ...389D001 Rev G 74 of 94 Auxiliary Blower Switch Status Wiring Option...
Page 75: ...389D001 Rev G 75 of 94 Auxiliary Cabinet Monitoring Wiring Option...
Page 76: ...389D001 Rev G 76 of 94 FlexAIR Canopy Exhaust Connection Wiring...
Page 77: ...389D001 Rev G 77 of 94...
Page 78: ...389D001 Rev G 78 of 94...
Page 79: ...389D001 Rev G 79 of 94...
Page 80: ...389D001 Rev G 80 of 94...
Page 81: ...389D001 Rev G 81 of 94...
Page 82: ...389D001 Rev G 82 of 94...
Page 83: ...389D001 Rev G 83 of 94...
Page 84: ...389D001 Rev G 84 of 94...
Page 85: ...389D001 Rev G 85 of 94...
Page 86: ...389D001 Rev G 86 of 94...
Page 87: ...389D001 Rev G 87 of 94...
Page 88: ...389D001 Rev G 88 of 94...
Page 89: ...389D001 Rev G 89 of 94...
Page 90: ...389D001 Rev G 90 of 94...
Page 91: ...389D001 Rev G 91 of 94...
Page 92: ...389D001 Rev G 92 of 94...
Page 93: ...389D001 Rev G 93 of 94...