389D001 Rev G
41 of 94
Press & Hold the Up arrow pushbutton until the “Setup” menu is displayed ~ 3 seconds
Set Point One Configuration:
- Press the Down arrow pushbutton to highlight [Set 1]
- Press the Right arrow pushbutton to access the Set 1 setting list
- Use the UP and Down arrow pushbuttons to highlight “db”
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
- Press the Down arrow pushbutton to highlight [1 act]
- Press the Right arrow pushbutton to access the 1 act setting list
- Use the Up and Down arrow pushbuttons to highlight “dir”.
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
- Press the Down arrow pushbutton to highlight [SP1 D]
- If this is not at ‘0’ press the Right arrow pushbutton to access the SP1 D setting list
- Set the delay to ‘0’ using Right and Left pushbuttons to select the digit and Up and Down pushbuttons
increase/decrease the value.
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
Set Point Two Configuration:
- Press the Down arrow pushbutton to highlight [Set 2]
- Press the Right arrow pushbutton to access the Set 2 setting list
- Use the UP and Down arrow pushbuttons to highlight “db”
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
- Press the Down arrow pushbutton to highlight [2 act]
- Press the Right arrow pushbutton to access the 2 act setting list
- Use the Up and Down arrow pushbuttons to highlight “rev”.
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
- Press the Down arrow pushbutton to highlight [SP2 D]
- If this is not at ‘0’ press the Right arrow pushbutton to access the SP2 D setting list
- Set the delay to ‘0’ using Right and Left pushbuttons to select the digit and Up and Down pushbuttons
to increase/decrease the value.
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
Set Point One Value and Deadband settings:
- Press & Hold the Up arrow pushbutton until the “Main” menu is displayed
- Press the Up and/or Down arrow pushbuttons to highlight “SP1”
- Press the Right arrow pushbutton to access SP1 value setting
- Use the Right and Left pushbuttons to select the digit and Up and Down pushbuttons increase/decrease the
value.
This is the “high” or “upper” setpoint. Set value to +0.10 above cabinet set point pressure
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
- Press the Down arrow pushbutton to highlight “SP 1 db
- Press the Right arrow pushbutton to access the SP1 db value setting
- Use the Right and Left pushbuttons to select the digit and Up and Down pushbuttons increase/decrease the
value.
This is the “high” or “upper” setpoint deadband. Set to 0.01
- Press & Hold the Right arrow pushbutton ~ 1 second then release to accept this setting
Summary of Contents for SterilGARD SG404
Page 60: ...389D001 Rev G 60 of 94...
Page 74: ...389D001 Rev G 74 of 94 Auxiliary Blower Switch Status Wiring Option...
Page 75: ...389D001 Rev G 75 of 94 Auxiliary Cabinet Monitoring Wiring Option...
Page 76: ...389D001 Rev G 76 of 94 FlexAIR Canopy Exhaust Connection Wiring...
Page 77: ...389D001 Rev G 77 of 94...
Page 78: ...389D001 Rev G 78 of 94...
Page 79: ...389D001 Rev G 79 of 94...
Page 80: ...389D001 Rev G 80 of 94...
Page 81: ...389D001 Rev G 81 of 94...
Page 82: ...389D001 Rev G 82 of 94...
Page 83: ...389D001 Rev G 83 of 94...
Page 84: ...389D001 Rev G 84 of 94...
Page 85: ...389D001 Rev G 85 of 94...
Page 86: ...389D001 Rev G 86 of 94...
Page 87: ...389D001 Rev G 87 of 94...
Page 88: ...389D001 Rev G 88 of 94...
Page 89: ...389D001 Rev G 89 of 94...
Page 90: ...389D001 Rev G 90 of 94...
Page 91: ...389D001 Rev G 91 of 94...
Page 92: ...389D001 Rev G 92 of 94...
Page 93: ...389D001 Rev G 93 of 94...