389D001 Rev G
45 of 94
Ladder Schematics
CHECKED
DRAWN
DATE
DATE
TITLE
THIS DRAWING AND ITS CONTENTS ARE THE
PROPERTY OF THE BAKER COMPANY. THE
INFORMATION CONTAINED HERE ON MAY NOT BE
USED OR COPIED IN ANY MANNER WITHOUT THE
WRITTEN PERMISSION OF THE BAKER COMPANY.
389E500
T
HE
B
AKER
C
OMPANY
161 Gatehouse Rd., Sanford, Maine 04073 USA
“Creating Immaculate Atmospheres”
W.Day
11/14/12
T.Martin
11/14/12
LADDER SCHEMATIC
Protective conductor terminal
!
DENOTES WIRE NO. 54, 22 A.W.G.
ALL WIRES 16 A.W.G. UNLESS OTHERWISE NOTED
DENOTES CANOPY TERMINAL BLOCK, TERMINALS 3 OR 4
LEGEND
Caution(Refer to
Accompanying documents)
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
A
ORIGINAL RELEASE
DATE
DRN
APP-BY
APP-DATE
B
SGX04
FOR
DWG NO.
TB-3,4
22
54 BL
SHEET
B
UPDATE JUMPER AND MSC BOARD
CONFIGURATION TO 40702 REV F, PER
ECR#01451
12/30/14
W.Day
APP-BY
APP-DATE
1 OF 2
XFMR
T1
51
61
120V
60
50
TB-4,5,6
5
6
1
2
11 12
7
8
10V
WT
18
WT
18
BK
18
BK
18
S.S.R 1
B1 B2
A1 A2
D2
B
D
C
COM
A
D1
C2
C1
TB-1,2,3
LINE
GN/YW
12
1
BK
12
2
BK
12
NEUTRAL GROUND
0
TB-1,2,3
TB-4,5,6
WT
12
CB1 20A
BLOWER
OUTLET
LIGHT
UV LIGHT
MUTE
COMMON
POWER REQUIREMENT
115VAC 20A. 60HZ. 1Ø
100VAC 20A. 50/60Hz. 1Ø
MAXIMUM CURRENT 16 A.
.2A
GN/
Y
W
#1 #2
B
B
OPTIONAL LIFT
FEMALE
REAR
VIEW
TO
PUMP
MOTOR
LIFT
SYSTEM
XFMR
GN/
YW
GN/
YW
BK
,1
BK
,2
J2
-3
J2
-2
WT
70
BK
71
J2-
1
TB-1,2,3
TB-4,5,6
C
NO
PROBE
OPTIONAL
AFM
OPTIONAL UV
OR 16
33
UVL
WT
18
30
U.V. BALLAST
WT
BL
J1-4
34
33
YW
18
34
34
RD 16
RD
18
BK
18
B1 B2
32
BK
18
31
TB-1,2,3
.61A
J1-1
BL
18
33
33
YW
18
34
RD
18
TB-4,5,6
YW
18
VIEWSCREEN
CLOSED
NC
PROX 2
18
GND
NO
J18
LT
UV
+12V
OUTLET
BLOWER
PROX 1
J6
J24
PROX 1
J7
1
J4
J3
LT
UV
+12V
OUTLET
BLOWER
BK
J2
2
J15
J12
PROX 2
PROX 1
GND
J16
GND
J1
BLOWER
STATUS
BN
BL
NO
54
J9
BN
RD
PROX 2 OUT
SPEAKER OUT
GND
J1
9
J5
NC
FAN ON
GND
BLOWER
ENABLE
J13
J2
1
NC
EXHAUST
REQUEST
J2
3
18
BK
NO
J10
56
J2
0
2
55
GND
PROX 2
1
53
+12V
+12V
1
RD
J2
SASH
ALARM
BLOWER
OVERIDE
BL
J8
J14A
J17
KEYPAD
ALARM
EX
T
.
IN
T
.
FAN ON
GND
C
C
C
RD
BK
READY-SAFE CABLE
TO (SASH) ON
MOTOR CONTROL
GND
FLEX AIR
J31
+12V
ALM
RD
RD
12
OR
15
RD
14
WT
WT
14
14
WT
12
AC
AC
W
U
V
EARTH
22
0V
11
5V
GN
12
J1-7
MOTOR
BL
GN
13
J1-6
J1-3
BLOWER
BK
14
J1-5
BK
12
GN/YW
12
WT
12
10
BK
SASH
TB-1,2,3
BK
12
11
12
BK
12
C1 C2
READY-SAFE CABLE
TO J14 ON
CONTROL BOARD
GN
16
WT
16
BK
16
WT
16
WT
16
BK
16
BK
16
BK
16
SELF-RESETTING BREAKER
J1-8
J1-9
TB-1,2,3
40
40
40
43
43
J1-2
LOAD
44
RECEPTACLE
LINE
P
CB2 5A 2
A1 A2
BK
16
42
41
TB-4,5,6
BL18
YW
TB-4,5,6
LH
1
LH
2
TB-1,2,3
21
T5,FL
BL
RD18
3
BK
2
FEMALE
REAR
VIEW
BL18
1
3
BK
18
1
2
MALE
REAR
VIEW
25
BL
J2-1
J2-2
J2-3
FL.
BALLAST
RD
D1 D2
YW
WT
GN/YW
18
GN
RD18
26
RD
TB-4,5,6
23
24
BK,#2
BK,#2
BK,#1
BK,#1
GN/YW
18
VIEWSCREEN
SAFE LEVEL
OPTIONAL CONFIGURATIONS
FROM 0 TO 4 OUTLETS
GND
PRES
J32
+12V
SHIELD
TB-4,5,6
Summary of Contents for SterilGARD SG404
Page 60: ...389D001 Rev G 60 of 94...
Page 74: ...389D001 Rev G 74 of 94 Auxiliary Blower Switch Status Wiring Option...
Page 75: ...389D001 Rev G 75 of 94 Auxiliary Cabinet Monitoring Wiring Option...
Page 76: ...389D001 Rev G 76 of 94 FlexAIR Canopy Exhaust Connection Wiring...
Page 77: ...389D001 Rev G 77 of 94...
Page 78: ...389D001 Rev G 78 of 94...
Page 79: ...389D001 Rev G 79 of 94...
Page 80: ...389D001 Rev G 80 of 94...
Page 81: ...389D001 Rev G 81 of 94...
Page 82: ...389D001 Rev G 82 of 94...
Page 83: ...389D001 Rev G 83 of 94...
Page 84: ...389D001 Rev G 84 of 94...
Page 85: ...389D001 Rev G 85 of 94...
Page 86: ...389D001 Rev G 86 of 94...
Page 87: ...389D001 Rev G 87 of 94...
Page 88: ...389D001 Rev G 88 of 94...
Page 89: ...389D001 Rev G 89 of 94...
Page 90: ...389D001 Rev G 90 of 94...
Page 91: ...389D001 Rev G 91 of 94...
Page 92: ...389D001 Rev G 92 of 94...
Page 93: ...389D001 Rev G 93 of 94...