389D001 Rev G
18 of 94
SPEED CONTROLLER ADJUSTMENTS
:
If a speed controller adjustment is necessary;
Step 1:
Remove thumbscrews or shipping bolts from front fascia panel and raise, engaging the left and right
panel supports.
Step 1:
Using a ¼” nut driver or ratchet wrench, remove the aluminum cover from the
electrical mounting board located on the left hand side. Set the left DIP switch down to
take out of Automatic mode. A green LED will ‘flash’ indicating this manual setting.