Step 6
: Loosen the (4) 5/16” hex nuts, (2 on each side)
of the electrical board. Do not completely remove
these.
Step 7
: Create some slack in the black supply power
cable by pulling down. (A ladder may be required to
disconnect cable ties on top of the cabinet).
Step 8
: Lower the electrical mounting onto the lower two studs by lifting up and out. You must
also move the female end of the blower electrical plug (left side) disconnected earlier and insert in
the cutout provided on electrical board.