PanaFlow Z3 User’s Manual
59
Chapter 4. Error Codes and Troubleshooting
4.10 Option I/O Errors
Table 23: Option I/O Errors Description
Error Code
Error Message
Description
A1:AnalogCh(S2:3) Error!
ADC Channel(S2:3) is not
responding
Analog input /RTD input is not working. If error
persists after power cycle, contact BakerHughes
factory
A2:AnalogCh (S2:4) Error!
ADC Channel(S2:4) is not
responding
Analog input /RTD is not working. If error persists
after power cycle, contact BakerHughes factory
A3:AnalogCh (S2:1) Error!
DAQ Channel (S2:1) is not
responding
Analog output (4-20mA) is not working. If error
persists after power cycle, contact BakerHughes
factory
A4:AnalogCh (S2:2) Error!
DAQ Channel (S2:2) is not
responding
Analog output (4-20mA) is not working. If error
persists after power cycle, contact BakerHughes
factory
A6:(S2:3)Ch Not
Calibrated
Error occurs when Analog
Input/RTD(S2:3) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BakerHughes
factory
A7:(S2:4)Ch Not
Calibrated
Error occurs when Analog
Input/RTD (S2:4) are not
calibrated
Calibrate the Analog Input/RTD input. If error
persists after calibration, contact BakerHughes
factory
A10:(S2:3)Input
NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not
connected at Channel (S2:3).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C at
Channel (S2:3)
Check connectivity for Analog Input/RTD input and
RTD temperature. Try power cycling the meter. If
error persists after power cycle, contact
BakerHughes factory
A11:(S2:4)Input
NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not
connected at Channel (S2:4).
RTD Input: Error occurs when
RTD input is not connected or
temp greater than 390 deg C at
Channel (S2:4)
Check connectivity for Analog Input/RTD input and
RTD temperature. Try power cycling the meter. If
error persists after power cycle, contact
BakerHughes factory
A12:(S2:3)Ch OverRange
Err!
Exceeds input values. For
analog input (S2:3) greater
than 21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after power
cycle, contact BakerHughes factory
A13:(S2:4)Ch OverRange
Err!
Analog input(S2:4) greater than
21mA
Ensure analog input current less than 21mA. Try
power cycling the meter. If error persists after power
cycle, contact BakerHughes factory
A24:Aout(S2:1)OutOfRang
e!
When output from analog
output(S2:1) exceeds 21 mA or
less than 3.6 mA
Check the flow velocity. If velocity is within limits and
error still persists, contact BakerHughes factory
A25:Aout(S2:2)OutOfRan
ge!
When output from analog
output(S2:2) exceeds 21 mA or
less than 3.6 mA
Check the flow velocity. If velocity is within limits and
error still persists, contact BakerHughes factory
A31:(S2:3)Ch
UnderRange!
Lesser input values. For analog
input(S2:3) between 3.6 mA to
0.25mA
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BakerHughes factory
A32:(S2:4)Ch
UnderRange!
Lesser input values. For analog
input(S2:4) between 3.6 mA to
0.25mA.
Check input analog current is between 3.6 mA to
21mA. If error persists, contact BakerHughes factory
Summary of Contents for Panametrics PanaFlow Z3
Page 1: ...Flow 910 311 Rev C April 2019 Panametrics com PanaFlow Z3 User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 8: ...Contents vi PanaFlow Z3 User s Manual...
Page 12: ...Preface x PanaFlow Z3 User s Manual no content intended for this page...
Page 21: ...PanaFlow Z3 User s Manual 9 Chapter 2 Installation Figure 8 Lifting the PanaFlow Z3 System...
Page 25: ...PanaFlow Z3 User s Manual 13 Chapter 2 Installation Figure 12 Remote Mount Wiring Diagram...
Page 26: ...Chapter 2 Installation 14 PanaFlow Z3 User s Manual no content intended for this page...
Page 58: ...Chapter 3 Programming 46 PanaFlow Z3 User s Manual no content intended for this page...
Page 86: ...Appendix A Specifications and Model Configurations 74 PanaFlow Z3 User s Manual...
Page 104: ...Appendix B Data Records 92 PanaFlow Z3 User s Manual no content intended for this page...
Page 108: ...Appendix D CE Mark Compliance 96 PanaFlow Z3 User s Manual no content intended for this page...
Page 110: ...Warranty 98 PanaFlow Z3 User s Manual no content intended for this page...
Page 116: ......