Table 8: Available measurement parameters
Option Bar
Description
Good
Bad
VEL
Displays the flow velocity.
N.A.
N.A.
VOLUM
Displays the volumetric flow.
N.A.
N.A.
+TOTL
Displays the forward totalized volume flow.
N.A.
N.A.
-TOTL
Displays the reverse totalized volume flow.
N.A.
N.A.
TIMER
Displays the total flow measurement time.
N.A.
N.A.t
MDOT
Displays the mass flow.
N.A.
N.A.
+MASS
Displays the forward totalized mass flow.
N.A.
N.A.
-MASS
Displays the reverse totalized mass flow.
N.A.
N.A.
SS up
Displays the signal strength for the upstream direction.
50 - 75
<50 or >75
SS do
Displays the signal strength for the downstream direction.
50 - 75
<50 or >75
SNDSP
Displays the measured speedd of sound in the gas.
N.A.
N.A.
Tup
Displays the upstream ultrasonic signal transit time.
N.A.
N.A.
Tdown
Displays the downstream ultrasonic signal transit time.
N.A.
N.A.
DELTA
Displays the transit time difference between the upstream and
downstream signals.
N.A.
N.A.
Tot K
Displays the total K factor.
N.A.
N.A.
PEAK%
Displays the percentage of peak (set to +50 by default).
N.A.
N.A.
Qup
Displays the signal quality for the upstream direction.
≥ 1200
-400 to +400
Qdown
Displays the signal quality for the downstream direction.
≥ 1200
-400 to +400
AMPup
Displays the value for the signal amplitude of the upstream direction.
24 ± 5
<19 or >29
AMPdn
Displays the value for the signal amplitude of the downstream direction.
24 ± 5
<19 or >29
CNTup
Displays the AGC DAC count for the upstream gain setting.
N.A.
N.A.
CNTdn
Displays the AGC DAC count for downstream gain setting.
N.A.
N.A.
P#up
Displays signal peaks for the upstream direction.
100 - 2300
<100 or >2300
P#dn
Displays signal peaks for the downstream direction.
100 - 2300
<100 or >2300
TEMP
Displays the gas temperature (from 0/4-20mA input).
N.A.
N.A.
PRESR
Displays the gas pressure (from 0/4-20mA input).
N.A.
N.A.
AcVOL
Displays the actual volumetric flow.
N.A.
N.A.
StVOL
Displays standard volumetric flow.
N.A.
N.A.
Tu S
1
Displays Skan transit time upstream.
N.A.
N.A.
Td S
1
Displays Skan transit time downstream.
N.A.
N.A.
DT S
1
Displays Skan Delta T.
N.A.
N.A.
Tu M
1
Displays Measure transit time upstream.
N.A.
N.A.
Td M
1
Displays Measure transit time downstream.
N.A.
N.A.
DT M
1
Displays Measure Delta T.
N.A.
N.A.
Vinst
Displays the instantaneous velocity.
N.A.
N.A.
1
availabloe only if Burst Mode = S/M
25
Summary of Contents for PanaFlow Z1G
Page 1: ...PanaFlow Z1G Z2G User s manual 910 321 Rev A...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 9: ...no content intended for this page 1...
Page 21: ...Figure 12 Remote mount electronics transducer and preamplifier wiring ref dwg 702 731 732 13...
Page 28: ...no content intended for this page 20...
Page 30: ...no content intended for this page 22...
Page 38: ...no content intended for this page 30...
Page 40: ...no content intended for this page 32...
Page 43: ...Table 13 Service record cont Date Description of service Performed by 35...