Reverse Current
a. Enter Advance Calibration mode by pressing both LO/SELECT and HI/CAL buttons at same time
b. Press HI/NEXT to access CAL mode
c. Press HI/NEXT to get to Reverse Current mode HICC
d. Press LO/SELECT to confirm.
e. Press HI/SELECT to navigate to reverse option.
f. Wait while mA reverse option is processing.
g. GREEN LED is 20mA RED LED is 4mA
h. When it completes. Press any button to return to main menu
See Visual Below for Reverse Calibration.