CALIBRATION MODE 

 
  The Calibration Mode allows the user to reprogram the 
3000+ Recorder to different parameters, calibrate the 
Zero and Span and assign and set the relay outputs. 
 
  To enter the Calibration Mode, press the MENU key 
and the following display screen will appear: 
 

 

PRESS UP FOR CALIBRATION 

PRESS DOWN FOR STATUS 

9

 

 
 
  If you decide not to enter the Calibration Mode, press 
the MENU key and the display will return to the normal 
operating screen. 
 
  When in the Calibration Mode and the desired change 
has been made and no other changes are desired, you may 
skip to the end of the calibration screens by pressing the 
MENU key. 
 
  To enter the Calibration Mode, press the UP arrow key 
and the following screen will appear: 
 

 

SECURITY ID 

INPUT 4 DIGIT ID 0000 

 
 
  This screen is the security identification which requires 
a security code to be entered before access is allowed into 
the Calibration Mode. This code is set to 0000 at the 
factory. You may set a new code when in the Calibration 
screen. 
 
  To enter a security code use the RIGHT arrow key to 
move the cursor under the digit to be changed and use the 
UP or DOWN arrow keys to change the digit to the 
desired value. The values can be from 0 to 9 and A to F. 
 
  After entering the security code, press the ENTER key. 
If the code is incorrect the display will return to the 
normal operating screen. If the code is correct, the 
following screen will appear: 
 

 

MEASUREMENTS UNITS 

UNIT INITIALS=GPM 

 
 
  This screen allows the user to select the engineering 
unit of measure for the flow rate. To change the units, 
press the UP or DOWN arrow keys until the desired unit 
is shown. The engineering units available are: 
 
 
 

 

 
  GPM  = Gallons per Minute 
  GPD  = Gallons per Day 
  MGD  = Million Gallons per Day 
  CFS  = Cubic Feet per Second 
  CFM  = Cubic Feet per Minute 
  CFH  = Cubic Feet per Hour 
  CMM = cubic Meters per Minute 
  OPTION  =  Any three letter label can be entered 
 
  After you have selected the unit desired, press the 
ENTER key. If you have selected OPTION, skip to the 
section entitled OPTION on Page 14. 
 
 The next screen, after selecting a standard 
measurement, allows the selection of the multiplier for the 
unit of measurement and will appear as follows: 
 

 

UNIT MULTIPLIER 

GPM  X  1,000 

 
 
  Press the UP or DOWN arrow key to select the desired 
flow multiplier. The following multipliers are available: 
 
 X.001 
 X.01 
 X.1 
 X1 
 X10 
 X100 
 X1,000 
 X10,000 
 
  Press the ENTER key to store the multiplier. The 
following screen will appear: 
 

 

FULLSCALE UNITS=GPM 

FULLSCALE X  1,000 

 
 
  This screen allows the setting of the full scale range. 
Press the UP or DOWN arrow keys to change each digit 
to the desired value and the RIGHT arrow key to move 
the cursor under each digit. Press the ENTER key and the 
following screen will appear: 
 

 

TOTALIZER MULTIPLIER 

X  1,000 

 
  This screen allows the selection of the totalizer 
multiplier. Press the UP or DOWN arrow keys to select 
the desired multiplier and then the ENTER key. The 
following multipliers are available: 

Summary of Contents for 3000+

Page 1: ...ng Diagram 20 UNPACKING INSPECTION To avoid damage in transit Eastech Badger products are shipped to the customer in special shipping containers Upon receipt of the product perform the following unpac...

Page 2: ...or pressure or temperature totalization alarm indication and is used to program the recorder through menu driven prompts RECORDING 12 inch circular chart with 24 hour 7 day or 31 day selectable chart...

Page 3: ...a window The door is provided with two lockable latches Caution should be used when mounting the enclosure so that it is not exposed to direct sunlight A sun shield is recommended to prevent the temp...

Page 4: ...e dimensions If the outdoor enclosure has the window option do not mount facing the sun or install a sun shade to prevent overheatingoftheelectronicsinsidetheenclosure WIRING CONNECTIONS 4 20 MA OUTPU...

Page 5: ...en the screws on the connectors Strip off about 1 4 inch of insulation from the wires Insert the wire into the proper connector and tighten the screw securely After connecting all of the wires to the...

Page 6: ...ow not differential DC power must be connected in series with the load as shown above DRY CONTACT RELAYS MAX LOAD 250 MA 4 20 MA OUTPUT NO C SW 1 NC NO C SW 2 NC 1 2 3 4 5 6 7 8 TB5 RELAY OUTPUTS Rela...

Page 7: ...reen informs the user of the Model of the recorder This screen will be displayed for a few seconds The next screen to be displayed is the following SOFTWARE REVISION X XX SERIAL NUMBER 12345678 This s...

Page 8: ...tion of this manual CHT This indicates that the chart has made a complete rotation and the chart needs to be changed INT Indicates a contact integrator pulse output has occurred EE EEPROM read write e...

Page 9: ...t the following screen will appear MEASUREMENTS UNITS UNIT INITIALS GPM This screen allows the user to select the engineering unit of measure for the flow rate To change the units press the UP or DOWN...

Page 10: ...1 55 POWER 1 58 POWER 1 95 POWER 1 2 POWER 3 2 POWER 5 2 POWER OPTION The linear function is used where the transmitter output is proportional to flow The 1 55 power is used with 1 2 3 24 Parshall flu...

Page 11: ...to be at the span value if the UP or DOWN arrow keys are not pressed When pressing the UP or DOWN arrow keys the display will indicate the direction of the pen movement Once the pen is positioned to s...

Page 12: ...the channels to be summed must be input to the recorder when making this adjustment For example if the input signal is 4 20 mA for 0 to 100 then 4 mA must be input to all channels to be summed to mak...

Page 13: ...t points If the High alarm is set for 60 percent and the hysteresis is 3 percent then the High alarm will trip closed at 63 percent and open at 57 percent Press the UP or DOWN arrow keys to the desire...

Page 14: ...or DOWN arrow keys to select YES or NO and press the ENTER key The following screen will appear UNIT MULTIPLIER AAA X 1 000 This screen allows the selection of the desired flow rate multiplier Press...

Page 15: ...reens appear during the chart pen adjusting procedure one of the following problems probably occurred 1 There was not a 4 20 mADC analog signal to the recorder analog channel you selected while adjust...

Page 16: ...N OF DECIMAL WHOLE NUMBER TO HEXADECIMAL NUMBER 1 Divide the decimal number by 16 Example 57420 16 3588 75 2 Multiply only the fractional part of the product by 16 to arrive at the first character in...

Page 17: ...lue of the next character Example 224 0 224 3 Multiply the previously found product by 16 Example 224 X 16 3584 4 Repeat steps 2 and 3 in a cyclical fashion until you have added the last hexadecimal c...

Page 18: ...18 INSERT DWG 600662 9999 3000 RECORDER ASSY DWG...

Page 19: ...15 SPACER 6 EA 2 0 34 400006 0025 SET SCREW 4 40 X 3 16 EA 1 0 35 524720 CHART CLAMP BODY EA 1 0 36 400006 0071 SET SCREW 5 40 X 5 16 EA 1 0 37 512610 COVER W LOGO EA 1 0 38 524743 CHART KNOB EA 1 0 3...

Page 20: ...that pen arm transport tie down is removed from unit Increase damping See IOM page 10 Ok Now CONTACT FACTORY Will pen arm move with input now CONTACT FACTORY YES GO TO NEXT PAGE Do pen arm display ag...

Page 21: ...Loop Check chart rotation speed chart knob tightness power outages motor backlash Chart turn ok CONTACT FACTORY Chart turn ok YES NO Fast Slow Not at all NO YES YES 3000 should be operating properly...

Page 22: ...CULAR PURPOSE Eastech Badger shall not be liable for any defects attributable to acts or omissions of others after shipment nor any consequential incidental or contingent damage whatsoever NUCLEAR DIS...

Reviews: