background image

 

-TOTALIZER WORD- 

0

15

 

000  0000  0000 

 
  The Totalizer Word is required to setup the time base to 
properly totalizer in the selected option units. The 
totalizer word is defined as follows: 
 

 

HEX

  

  

2,000,000

  x  

ec

in Units/S

 

Flow

 

Max.

Multiplier

Totalizer 

=

 

 
EXAMPLE: 
Maximum flow  =  1800  x  1 BPM (Barrels/min) 
Totalizer multiplier  =  BAR  x  100 
Totalizer Word  = 
[100/(1800/60)] x 2,000,000  =  6666666.6 
 

Converted to Hexadecimal  =  65B9AA (see the Decimal 
to Hexadecimal Conversion Section). 
 
The TOTALIZER WORD is 0000 0065 B9AA 

 
NOTE: The totalizer word must be greater than Hex 
1000. 

 
  Use the UP and DOWN arrow keys to change each 
digit to the required value and the RIGHT arrow to move 
the cursor. The ENTER key must be depressed after each 
set of four digits. Return to Page 10 where the next screen 
will be the Chart Speed selection. 
 
 
 

 
 
 
 
 
 
 

ERROR SCREENS 

 
  There are several Error Screens that will be displayed 
by the recorder if the microprocessor detects errors in the 
values programmed into the recorder. 
 

 

-CHART PEN ZERO- 

<INPUT ZERO ERROR> 

 
 

OR 

 
 

-CHART PEN SPAN- 

<INPUT SPAN ERROR> 

 
 
  If either of the two above Error Screens appear during 
the chart pen adjusting procedure, one of the following 
problems probably occurred: 
 
  1.  There was not a 4-20 mADC analog signal to the 
recorder analog channel you selected while adjusting the 
chart zero span. 
 
  2.  An analog input signal channel was selected 
without adjusting the chart pen zero and span of this 
channel. 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
  3.  The Zero and Span of the recorder have been 
adjusted too close to the same input value. 
 
  To correct these errors, readjust the Chart Pen Zero and 
Span while inputting the 4-20 mADC analog signal to the 
proper selected recorder channel. Refer to the Calibration 
Mode Section of this manual. 
 

 

-STORING PARAMETERS- 

<TOTALIZER WORD ERROR> 

 
 
  If the above Error Screen appears after pressing the 
ENTER key to store the calibration adjustments, this 
indicates that you have selected a totalizer multiplier that 
is too low for your selected flow rate. To correct this 
error, re-enter the Calibration Mode and when the 
Totalizer Multiplier Screen appears, increase the totalizer 
multiplier and press the ENTER key. Then press the 
MENU key to store the new multiplier and return to the 
normal operating screen. 
 
  If you still have a Totalizer Word Error message in the 
Storing Parameters screen, repeat the above procedure 
until the error is not present. 
 

Summary of Contents for 3000+

Page 1: ...ng Diagram 20 UNPACKING INSPECTION To avoid damage in transit Eastech Badger products are shipped to the customer in special shipping containers Upon receipt of the product perform the following unpac...

Page 2: ...or pressure or temperature totalization alarm indication and is used to program the recorder through menu driven prompts RECORDING 12 inch circular chart with 24 hour 7 day or 31 day selectable chart...

Page 3: ...a window The door is provided with two lockable latches Caution should be used when mounting the enclosure so that it is not exposed to direct sunlight A sun shield is recommended to prevent the temp...

Page 4: ...e dimensions If the outdoor enclosure has the window option do not mount facing the sun or install a sun shade to prevent overheatingoftheelectronicsinsidetheenclosure WIRING CONNECTIONS 4 20 MA OUTPU...

Page 5: ...en the screws on the connectors Strip off about 1 4 inch of insulation from the wires Insert the wire into the proper connector and tighten the screw securely After connecting all of the wires to the...

Page 6: ...ow not differential DC power must be connected in series with the load as shown above DRY CONTACT RELAYS MAX LOAD 250 MA 4 20 MA OUTPUT NO C SW 1 NC NO C SW 2 NC 1 2 3 4 5 6 7 8 TB5 RELAY OUTPUTS Rela...

Page 7: ...reen informs the user of the Model of the recorder This screen will be displayed for a few seconds The next screen to be displayed is the following SOFTWARE REVISION X XX SERIAL NUMBER 12345678 This s...

Page 8: ...tion of this manual CHT This indicates that the chart has made a complete rotation and the chart needs to be changed INT Indicates a contact integrator pulse output has occurred EE EEPROM read write e...

Page 9: ...t the following screen will appear MEASUREMENTS UNITS UNIT INITIALS GPM This screen allows the user to select the engineering unit of measure for the flow rate To change the units press the UP or DOWN...

Page 10: ...1 55 POWER 1 58 POWER 1 95 POWER 1 2 POWER 3 2 POWER 5 2 POWER OPTION The linear function is used where the transmitter output is proportional to flow The 1 55 power is used with 1 2 3 24 Parshall flu...

Page 11: ...to be at the span value if the UP or DOWN arrow keys are not pressed When pressing the UP or DOWN arrow keys the display will indicate the direction of the pen movement Once the pen is positioned to s...

Page 12: ...the channels to be summed must be input to the recorder when making this adjustment For example if the input signal is 4 20 mA for 0 to 100 then 4 mA must be input to all channels to be summed to mak...

Page 13: ...t points If the High alarm is set for 60 percent and the hysteresis is 3 percent then the High alarm will trip closed at 63 percent and open at 57 percent Press the UP or DOWN arrow keys to the desire...

Page 14: ...or DOWN arrow keys to select YES or NO and press the ENTER key The following screen will appear UNIT MULTIPLIER AAA X 1 000 This screen allows the selection of the desired flow rate multiplier Press...

Page 15: ...reens appear during the chart pen adjusting procedure one of the following problems probably occurred 1 There was not a 4 20 mADC analog signal to the recorder analog channel you selected while adjust...

Page 16: ...N OF DECIMAL WHOLE NUMBER TO HEXADECIMAL NUMBER 1 Divide the decimal number by 16 Example 57420 16 3588 75 2 Multiply only the fractional part of the product by 16 to arrive at the first character in...

Page 17: ...lue of the next character Example 224 0 224 3 Multiply the previously found product by 16 Example 224 X 16 3584 4 Repeat steps 2 and 3 in a cyclical fashion until you have added the last hexadecimal c...

Page 18: ...18 INSERT DWG 600662 9999 3000 RECORDER ASSY DWG...

Page 19: ...15 SPACER 6 EA 2 0 34 400006 0025 SET SCREW 4 40 X 3 16 EA 1 0 35 524720 CHART CLAMP BODY EA 1 0 36 400006 0071 SET SCREW 5 40 X 5 16 EA 1 0 37 512610 COVER W LOGO EA 1 0 38 524743 CHART KNOB EA 1 0 3...

Page 20: ...that pen arm transport tie down is removed from unit Increase damping See IOM page 10 Ok Now CONTACT FACTORY Will pen arm move with input now CONTACT FACTORY YES GO TO NEXT PAGE Do pen arm display ag...

Page 21: ...Loop Check chart rotation speed chart knob tightness power outages motor backlash Chart turn ok CONTACT FACTORY Chart turn ok YES NO Fast Slow Not at all NO YES YES 3000 should be operating properly...

Page 22: ...CULAR PURPOSE Eastech Badger shall not be liable for any defects attributable to acts or omissions of others after shipment nor any consequential incidental or contingent damage whatsoever NUCLEAR DIS...

Reviews: