7-17
Operation & Maintenance Manual
Doc. Ref. MNGRM2CO2 REV. 3
Sensor 4mA Calibration
1.
On the GRM2 CO2 display/keypad, navigate to the
System Test
screen. This will halt the
sampling activity on the system for a few minutes.
NOTE: This is not required, but the sampling activity of the system may prove distracting whilst
performing the subsequent steps.
2.
Connect a CC01 programming/debug lead to the CC01 programming/debug port on the GRM2
CO2 network module PCB.
3.
Open a terminal program and connect to the CC01 using 9600 baud, 8 data bits, 1 stop bit, No
parity, No flow control.
4.
In the terminal program:
4.1.
Enter the ‘ver’ command to confirm successful communication has been established.
5.
On the Sensor PCB:
5.1. Press the
A
button to open the menu system.
5.2. Using the
C
and
D
buttons select menu option:
E:4
on the sensor PCB screen.
5.3. Press
B
.
6. In the terminal program:
6.1.
Enter the ‘adc’ command a few times and note the ADC 12 0: value reported
(highlighted below in yellow).
Channel Bits mVolts Value.
--------------------------------------
Pressure
187
456
1111
Detector
2
4
Source
5
12
ADC 3: 786
1918
DAC offset 518
1264
Amb. Temp. 210
512
307
ADC12 0: 400
ADC12 1: 2
ADC12 2: 1
7.
On the Sensor PCB:
7.1. Use the
C
and
D
buttons to adjust the 4mA output.
8.
Repeat steps 6 & 7 until the ‘adc’ command output reads 400 +/- 1.
9.
On the Sensor PCB:
9.1. Press
B
.
NOTE: Pressing the A button rather than the B button exits without performing the
calibration.
9.2. Press the
A
button to close the menu system.
NOTE: The Sensor 4mA calibration value will be displayed on exit.
Sensor 20mA Calibration
1.
On the GRM2 CO2 display/keypad, navigate to the
System Test
screen. This will halt the
sampling activity on the system for a few minutes.
NOTE: This is not required, but the sampling activity of the system may prove distracting whilst
performing the subsequent steps.
2.
Connect a CC01 programming/debug lead to the CC01 programming/debug port on the GRM2
CO2 network module PCB.
3.
Open a terminal program and connect to the CC01 using 9600 baud, 8 data bits, 1 stop bit, No
parity, No flow control.
Summary of Contents for Parasense GRM2 CO2 Series
Page 7: ...1 2 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 GENERAL ARRANGEMENT...
Page 14: ...2 7 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 NETWORK MODULE CONNECTIONS...
Page 74: ......
Page 75: ......
Page 76: ...2 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 GRM2 CO2 CONNECTION SCHEMATICS PS0748...
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ...3 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 GENERAL ARRANGEMENT PM2098...
Page 83: ......
Page 84: ...4 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 PANEL MOUNTING DETAILS PM2247...
Page 85: ......
Page 86: ...5 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 CONNECTION SCHEMATIC PS0753...
Page 87: ......
Page 88: ......
Page 89: ......
Page 91: ......
Page 93: ......
Page 95: ......
Page 96: ......