7-8
Operation & Maintenance Manual
Doc. Ref. MNGRM2CO2 REV. 3
REPLACING A SAMM BLOCK
Switch off the mains supply to the monitor
and wait for a few seconds while the power
supply discharges. Open the enclosure door
with the key provided and switch off at the
internal power input module using the red
button
(1)
.
Disconnect the hose from the faulty SAMM
block
(2)
leaving the white non-return valve
connected to the SAMM block. Unplug the
SAMM block wiring connector
(3)
.
Remove the four fixing screws
(4)
that hold
the faulty SAMM block to the base of the
enclosure, being careful not to damage the
gasket between the SAMM and the inside of
the enclosure.
To install the new SAMM block, fit the new SAMM block in the vacant position. (Note: It will only fit
with the 8 short sampling pipes facing towards the front of the enclosure.) Fit the shake-proof
washers and fixing screws and tighten. Re-attach the hose to the non-return valve of the new SAMM
block and plug in the SAMM block wiring connector.
If you are wiring the new SAMM block directly to the network module for pump 5-8, 9-12 or 13-16,
then you must connect the wiring for the new SAMM block pump marked 1 to the far left pair of
terminals on the associated network module connector and the wiring for pump 2 to the next pair of
terminals to the right and so on.
Switch on at the internal power input module using the green button. Close and lock the enclosure
door and reinstate the mains supply.
1
2
3
4
Summary of Contents for Parasense GRM2 CO2 Series
Page 7: ...1 2 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 GENERAL ARRANGEMENT...
Page 14: ...2 7 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 NETWORK MODULE CONNECTIONS...
Page 74: ......
Page 75: ......
Page 76: ...2 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 GRM2 CO2 CONNECTION SCHEMATICS PS0748...
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ...3 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 GENERAL ARRANGEMENT PM2098...
Page 83: ......
Page 84: ...4 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 PANEL MOUNTING DETAILS PM2247...
Page 85: ......
Page 86: ...5 Operation Maintenance Manual Doc Ref MNGRM2CO2 REV 3 ADC 104 CONNECTION SCHEMATIC PS0753...
Page 87: ......
Page 88: ......
Page 89: ......
Page 91: ......
Page 93: ......
Page 95: ......
Page 96: ......