Multi-Zone Gas Monitors
P/N: 3015-5074 Rev 11
61
B.3.13. Sensor Data Register 0x0016h (22 Dec) (R, 82 Byte s)
Variable
Type
Length
Description
Pressure
Float
4
Manifold Pressure is PSIA
P_Volts
Float
4
Pressure sensor output Voltage
Vacuum_P
Float
4
Pressure with all value closed and pump on in PSIA
Ambient_P
Float
4
Absolute Ambient Pressure in PSIA
Box_T
Float
4
Enclosure Temperature in Degrees C
Box_T_Volts
Float
4
Box temp sensor output voltage
Bench_T
Float
4
Optical bench temperature in Degrees C
Bench_T_Volts
Float
4
Bench temp sensor output voltage
Bench_Z_T
Float
4
Optical bench temp in degrees C at last zero interval
Box_Z_T
Float
4
Box temp in degrees C at last zeroing
PkPk_int
UI
2
Current peak to peak A/D counts from detector
PkPk
Float
4
Current peak to peak voltage from detector
Ave_PkPk
Float
4
Running average voltage from detector
Zero_PkPk
Float
4
Voltage measured at last zeroing
Noise
Float
4
Largest Change in running average
AU
Float
4
Current absorbance value
Ave_AU
Float
4
Running Average of absorbance level
Bench_PPM
Float
4
PPM in bench based on zone gas selection (uncorrected for
pressure and temperature)
STP_PPM
Float
4
PPM corrected to STP (1 atm, 25 deg C)
Bench_UML
Float
4
Micromoles/liter in bench (uncorrected)
Ambient_UML
Float
4
Micromoles/liter corrected to ambient pressure
B.3.14. Release Zone Hold Register 0x0017h (23 Dec) (W, 10 Bytes)
Variable
Type
Length
Description
Rel_Hold
*
*
See description of STATUS REGISTER
B.3.15. Hold Zone Register 0x0018h (23 Dec) (W, 10 Byte s)
Variable
Type
Length
Description
Zone_Hold
*
*
See description of STATUS REGISTER
B.3.16. MZ Hold Mode
The MZ monitor can be made to hold or “dwell” on a particular zone if necessary. The length of the hold
time is defined by the HOLD TIME parameter in the System Data Register.
Placing the MZ Monitor into hold mode:
1. Read the MZ Status Register (0x0011h)
2. Modify the content of the status register structure to change the MODE paramet er to zone hold
mode. Active zone parameter to the zone you wish to hold
3. Send this updated status register structure back to the MZ using PRESE T MULTIPLE REGIS TE R
COMMA ND to the HOLD ZONE RE GIS TE R (0x 0018h).
Releasing the Zone Hold
1. Read the MZ monitor Status Register (0x0011h)
Summary of Contents for MultiZone
Page 52: ...Multi Zone Gas Monitors 46 P N 3015 5074 Rev 11 ...
Page 58: ...Multi Zone Gas Monitors 52 P N 3015 5074 Rev 11 ...
Page 60: ...Multi Zone Gas Monitors 54 P N 3015 5074 Rev 11 ...
Page 71: ...Multi Zone Gas Monitors P N 3015 5074 Rev 11 65 APPENDIX C SYSTEM MENU MAP ...
Page 72: ...Multi Zone Gas Monitors 66 P N 3015 5074 Rev 11 ...
Page 73: ...Multi Zone Gas Monitors P N 3015 5074 Rev 11 67 APPENDIX D AGENCY APPROVALS ...
Page 74: ...Multi Zone Gas Monitors 68 P N 3015 5074 Rev 11 ...
Page 76: ...Multi Zone Gas Monitors 70 P N 3015 5074 Rev 11 ...
Page 78: ...Multi Zone Gas Monitors 72 P N 3015 5074 Rev 11 ...