46
User Manual v2.0
MAG
5.6 SEPARATE VERSION SENSOR CONNECTION
All operations on the electrical connections must be carried out if the instrument is disconnected
from the mains and/or from the battery!
All operations performed on the electrical junction box or the electrical components of the device
must be carried out by duly trained and specialized personnel.
The sensor is supplied with cables connected and resin coated in the separation box unless otherwise
requested by the customer.
The following diagrams, for the connection of the separation cables between sensor and converter, may
vary according to the type of converter; therefore, follow the indications given in the relevant documentation.
Info: If cables in the coil are used, always use the ferrules for the connections in the terminal board
and adequately insulate the shields.
Info: use only cables supplied by B METERS srl.
ATTENTION: After making all the necessary connections, make sure to properly tighten the cable
glands and the cover, in order to protect the insulation of the cables and the electrical/electronic
boards from humidity and water in case the converter is installed outdoors. Refer to the manual of
the relative converter for more information.
A
B
T
C
D
E
F
A
B
T
C
D
E
F
F
E
D
C
T
F
B
A
SCATOLA DI DERIVAZIONE
SENSORE
SCATOLA DI DERIVAZIONE
CONVERTITORE
SOLO CAVI A DOPPIA
SCHERMATURA C028
Fig. 68
JUNCTION BOX OF THE
SENSOR
ONLY DOUBLESHIELDED
CABLES C028
JUNCTION BOX OF THE
CONVERTER