The manufacturer has identified all the power cables and signal cables that have to be connected
to the terminal block of the junction box (in case of REMOTE version) by marking them with bands
(with letters) in order to ease the identification of the correct connection terminals (Fig. 57).
CAUTION: When all the necessary cables are connected, be sure to properly tighten the cable glands
(see 1, Fig. 57) and the cover (see 2, Fig. 57) so as to guarantee good moisture and water resistance
of the cables and electrical/electronic boards if the converter is installed outdoors.
Any unused cable glands (see 3, Fig. 57) must always be closed with the relevant blind rubber
sealings (see 4, Fig.57).
The used cable glands (see 1, Fig. 57) are suitable for multi-wire cables with a minimum diameter of
5 mm up to a maximum diameter of 12 mm (see 4, Fig. 57).
Note: The cables supplied by B METERS srl are not suitable for being run directly underground
and/or underwater, and must not have direct exposure to sunlight. The technician in charge must
therefore provide suitable protections such as: corrugated pipes, raceways, or shielded cables.
Note: It is recommended to keep the separation cables detached from power cables and/or
electromagnetic interference sources.